Bispectralindex
BIScompatibility
AllfeaturesofthismeasurementareavailablewithallBx50monitors.
Fordetailedinformationregardingmodule,monitor,andaccessorycompatibility,see
theCARESCAPEModularMonitorsSupplementalInformationManual.
BISsafetyprecautions
BISwarnings
WARNING
—DEFIBRILLATORPRECAUTIONS—Patientsignalinputs
labeledwiththeCFandBFsymbolswithpaddlesare
protectedagainstdamageresultingfromdebrillation
voltages.Toensureproperdebrillatorprotection,useonly
therecommendedcablesandleadwires.
WARNING
Makesurethattheelectrodes,sensorandconnectorsdonot
touchanyelectricallyconductivematerial,includingearth.
WARNING
Whenusinganelectrosurgeryunit,notethatthe
measurementcablesdonotincorporatemeanstoprotect
againstburnsincaseofadefectiveESUreturnelectrode.To
avoidburnsatthemonitormeasurementsites,ensurethe
following:
●PropercontactoftheESUreturnelectrodetothepatient.
●ESUreturnelectrodeneartheoperatingarea.
●Measurementelectrodes,leadwiresandprobesfarfrom
thesurgicalsiteandtheESUreturnelectrode.
WARNING
Donotautoclavethesignalprocessingunit(BISx)orthedigital
signalconverter(DSC).Donotopenitforanyreason.
WARNING
Thesensormustnotbelocatedbetweendebrillatorpads
whenadebrillatorisusedonapatientconnectedtoBIS
devices.