GE CARESCAPE - Preparing the sidestream setup with Dual CO2 module

GE CARESCAPE
548 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CO2withCAPNOSTATMainstream,CapnoFlexLF,andDualCO2modules
1.Nasalcannula(1a,non-intubatedpatients)orairwayadapter(1b,intubated
patients)
2.Patientsampleline(intubatedpatients)
3.Watertrap
4.12.7cm(5in)male/maleadaptertubing
5.Sidestreamadapter
6.Sensor(uprightposition)
7.Referencecell
8.Zerocell
9.Sidestreamadaptertubing
10.Sensorcable
11.Scavengingtube
12.Module
13.Sidestreaminletconnector
PreparingthesidestreamsetupwithDualCO2module
1.Connectthesidestreamadaptertubingtotheinletconnectoronthemodule.
2.ConnectthescavengingtubetotheexhaustconnectorifN
2
Oorvolatileagents
areused.
3.Attachthe12.7-cm(5-in)adaptertubingconnectortotheairwayadapter.
4.Securethewatertraptotheadaptertubing.
5.Connectthenasalcannulaorsamplelinetothewatertrap.
6.Snapthesidestreamadapterwithitstubingintothesensor.
7.Connectthesensorcabletothemodule.ThemessageWarmingupappearson
themonitorscreen.Waituntilitdisappears.
8.Turnonthepump.
2062971-001CARESCAPEModularMonitors313