Non-invasivebloodpressure
InitialNIBPcuffinationpressure
PDMPSMTRAM
Adult135mmHg(18kPa)170mmHg(22.7kPa)160mmHg(21.3kPa)
Child125mmHg(16.7kPa)150mmHg(20.0kPa)140mmHg(18.7kPa)
Infant100mmHg(13.3kPa)120mmHg(16.0kPa)110mmHg(14.7kPa)
NOTE
PDMandTRAM:WhentheAutoInitialInatesettingis
enabled,theinitialcuffinationpressuresaredependenton
theNIBPmoduleusedandselectedcuffsize.Theinitialtarget
pressurepresetcanbeadjustedifyoudesirealowerorhigher
initialtargetpressure.
SelectingtheinitialNIBPcuffinationpressure
NOTE
PDMandTRAMonly.
Youcandeterminethecuffinationpressureautomaticallybasedonthecuffsize.
1.SelecttheNIBPparameterwindow.
2.SelectAutoInitialInate.
SettingthetargetNIBPinationpressure
NOTE
PDMandTRAMonly.
YoucanmanuallychangethetargetinationpressurefortherstNIBPmeasurement.
1.SelecttheNIBPparameterwindow.
2.CheckthatAutoInitialInateisnotselected.
3.SelectavaluefromtheInit.Pressurelist.
Selectingthecuffinationlimits
NOTE
PSMonly.
Black-coloredAdult/Childcuffhosesandblue-coloredInfantcuffhosesare
automaticallydetectedbythemonitorandinationlimitsaresetaccordingly.
However,ifthecuffhosescannotbedetectedautomatically,youmustsettheination
limitsmanually.Youcanalsoselecttheinationlimitswhentheautomaticdetection
isworking.
1.SelecttheNIBPparameterwindow.
2.SelecttheSetuptab.
3.SelectInfant,Child,orAdultfromtheInationLimitslist.
NIBPvolumeanddisplaysettings
AdjustingtheNIBPmeasurementcompletiontonevolume
1.SelecttheNIBPparameterwindow.
2062971-001CARESCAPEModularMonitors227