Printing
2.SelecttheArrhythmiatab.
3.SelectPrintonAlarmforthearrhythmiasyouwouldliketoprint.
NOTE
Forarrhythmiaalarmwaveformprinting,theprintingwill
continueuntil20secondshaspassedfromtheclearanceof
thelastactivearrhythmiaalarm(e.g.,10secondssaveddata,
arrhythmiaalarmduration+20secondsdata).
Printingwaveformsforotherthanarrhythmiaalarms
Thefollowingarethedefaultsettingsforotherthanarrhythmiaalarmwaveform
printouts:
●Theprintdelayis10seconds.
●Theprintdurationis30seconds.
●Thewaveformspeedis25mm/s.
Otherthanarrhythmiaalarmsthatprintarethehigh/lowalarmsforthefollowing:HR,
Art/ABP/Fem/UACsys/dia/mean,SpO
2
,andST.
1.SelectMonitorSetup>Printing.
2.SelecttheWaveformstab.
3.ChooseavaluefromthePrintonAlarmlist:
●No:Noalarmwaveformsprintduringanalarmcondition.
●High:Alarmwaveformsprintduringhighpriorityalarmconditionsonly.
●All:Alarmwaveformsprintduringanyalarmcondition.
Settingtheprintdelay
NOTE
Alarmwaveforms(arrhythmiaandnon-arrhythmia)start
byprinting10secondsofthemostrecentlysaveddata,
regardlessoftheDelaysetting.
1.SelectMonitorSetup>Printing.
2.SelecttheWaveformstab.
3.SelectavaluefromtheDelaylist:
a.0s:Manualwaveformprintingstartswithrealtimedata.
b.10s:Manualwaveformprintingstartswith10secondsofthemostrecently
saveddatarst.Afterthat,therealtimedatabeginstoprint.
Settingtheprintduration
NOTE
Non-arrhythmiaalarmsprint30secondsofwaveforms
regardlessofthePrintLengthsetting.
NOTE
Anarrhythmiaalarmwaveformcontinuestoprintupto20
secondsafterthearrhythmiaresolvesoranewarrhythmia
hasbeenidentied.Thelengthoftheeventshalldetermine
thelengthoftheprintout.
1.SelectMonitorSetup>Printing.
2.SelecttheWaveformstab.
2062971-001CARESCAPEModularMonitors437