Temperature
Stoppingthetemperaturemeasurement
1.Selectthetemperatureparameterwindow.
2.SelectMeasurement>Off.
Temperaturepracticalities
Eachtemperaturelabelcanbechangedtoreectthetemperaturemeasurement
site.
Individualtemperaturesitescanbeturnedoff.
Tbloodisobtainedfromapulmonaryarterycatheter.
Thedifferencebetweentwotemperaturesitescanbecalculatedanddisplayed.
Thedualtemperaturecableallowsatwo-channelmeasurement.Thetemperature
cableswitchcanbesetto400or700dependingonthetypeofprobeused.Since
only400issupported,makesurethatyouturntheswitchtothisposition.
Thesignalinputisahigh-insulationporttoensurepatientsafetyandtoprotectthe
deviceduringdebrillationandelectrosurgery.
Themonitorautomaticallycalibratesthetemperaturemeasurementsatstartup:
every10minutesforPSM,andeveryminuteforPDMandTRAM.
Temperaturetroubleshooting
ProblemSolution
TemperaturemeasurementfailsCheckthattheprobeadapterisproperly
connectedtotheacquisitionmodule.
Checkthattheprobeisproperlyconnectedtothe
probeadapter.
Checkthatyouareusingthecorrectprobeforthe
anatomicallocationbeingmonitored
Useaprobethatiscompatiblewithyoursystem.
Tryusingaknowngoodprobeincasethesensor
isdamaged.
Checkthattheacquisitionmoduleisproperly
connectedtothemonitor.
Checkthepatientconnection.
Checkthattherearenottwoidentical
measurementmodulesinthesystem.
Contactserviceiftheproblemcontinues.
2062971-001CARESCAPEModularMonitors259