6
Startingandendingmonitoring
Monitorandmodulecompatibility
TRAMandTram-RacmoduleswiththeB850only.
Softwarepackagesandterminology
Theterminologyusedindifferentsoftwarepackagesvaries:inORandPACU,youstart
orresetacase,andinothersoftwarepackagesyouadmitordischargeapatient.In
addition,someothermenuselectionsmayalsodifferaccordingtothelicensesinuse.
Readallinstructionscarefully.
Abouttheuserdefaultsettings
Userdefaultsettingsmeanthosesettings(start-upmode,proleetc.)thattheuser
hassavedintothemonitortoreplacethefactorydefaultsettings.Themonitoruses
thesesettingswhenitisturnedonandafterapoweroffsituationthatlastsmorethan
15minutes.Iftherearenouserdefaultsettings,factorydefaultsettingsareused.
InvasivepressurelabelsandPDMorTRAM
WARNING
WhenconnectingPDMorTRAM,theloadedIPlabelsmay
affectthechannellabelingofotheralreadyconnected
channels,andconsequentlyalsothealarmlimits.
Whenmonitoringisstarted,storedinvasivepressurelabelsforzeroedchannelswill
betransferredfromthePDMorTRAMtothemonitor.Labelsforthosechannelsthat
havenotbeenzeroedwillbetheonesselectedatthemonitor.Notethatthecurrently
storedlabelsmaydifferfromdefaults.
TheIPlabelssavedtoPDMorTRAMaremappedaccordingtothistable:
Fromthemonitor
toPDMorTRAM
ArtART
ABPART
CVPCVP
2062971-001CARESCAPEModularMonitors115