Messages
Message
Location
ExplanationWhattodo
Wrongcable
param.
Youhaveconnecteda
MasimoSetcabletothePDM
Nellcor,oraNellcorOxiMax
cabletothePDMMasimo.
Checkandusethecorrect
cable.
Wrongcable.Use
MasimoSet
al.area,
param.
YouhaveconnectedaNellcor
OxiMaxcabletothePDM
Masimo.
UseaMasimoSetcable.
Wrongcable.Use
NellcorOxiMax
al.area,
param.
Youhaveconnecteda
MasimoSetcabletothePDM
Nellcor.
UseaNellcorOxiMaxcable.
MessagesrelatedtoNIBPmeasurement
Forinformationregardingalarmprioritiesandescalationtimes,seetheCARESCAPEModular
MonitorsSupplementalInformationManual.
Makesureyouarefamiliarwiththegenericlayoutofthescreen.Thiswillhelpyouidentify
whereonscreenthefollowingmessagesappear.Themessagelocationisindicatedwiththe
followingabbreviations:
al.area=alarmarea
param.=parameterwindow
report=reportview
wavef.=waveformarea
Message
Location
ExplanationWhattodo
Airleakage
param.
Loosecufforcuffhose.Checkthecuffandcuffhose.
Artifacts
param.
Measurementisdisturbedby
artifact(e.g.patientmovement,
shivering,deepbreathing,
markedarrhythmiaorirregular
beats).
Calmthepatientandretry.
Calibrated
param.
Channelcalibratedsuccessfully.Waituntilthemessage
disappearsbeforestartinga
measurement.
Calibrating
param.
Calibrationofachannelisin
progress.
Noactionrequired.
CalibrationswitchON
param.
Thecalibrationswitchison.Contactauthorizedservice
personnel.
Callservice
param.
Technicalfault.Contactauthorizedservice
personnel.
472CARESCAPEModularMonitors2062971-001