Messages
Message
Location
ExplanationWhattodo
●Callservice:Errorx
●wherex=0-99
●param.
0=RAMfailure
1=ROMchecksumfailure
2=+15Vfailure
3=-15Vfailure
6=ADCerror
7=Watchdogtimetooshort
8=Watchdogtimetoolong
9=Watchdogactivated
10=EEPROMchecksumerror
11=Autozerorangeexceeded
12=Communicationwatchdog
activated
13=Notinuse
14=Tooearlyautocyclestart
●
Contactauthorizedservice
personnel.
●CheckNIBP●al.area
Systolicand/ordiastolicresults
missing.
●Checkthepatientstatus.
●CheckNIBPcuffandhoses.
●Repeatthemeasurement.
●Controlmeasurement●al.areaPressurealarmlimitexceeded.●Allowmeasurementto
complete.
●Checkthepatientstatus.
●Cuffocclusion
●param.
Occlusionduringmeasurement
oroverpressuredcuff.
●Checkthecuff.
●Cuffloose
●param.
Loosecufforcuffhose.●Checkthecuffandcuffhose.
●Cuffoverpressure
●param.
NIBPcuffissqueezedduring
measurement.
●CheckNIBPcuffandhoses.
●Repeatthemeasurement.
●Incorrectin.limits
●param.
Adultorchildcuffisused,butthe
selectedinfantmoderestricts
theinationpressuretoolowto
beabletomeasuretheblood
pressure.
●AdjustNIBPsetup.
●Longmeas.time
●param.
Themeasurementtimeislong.
Thetriggeringvaluesvary
accordingtothemoduleand
inationlimitsinuse:
●PSM:>2minforadult/child,
75sto80sforinfant
●PDM:>2minforadult/child,
85sforinfant
●TRAM:<3minforadult/child,
90sforinfant
●Checkthepatientstatus.
●Checkthecuffandhose
connections.
●Restartthemeasurement.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●NIBPairleakage/
NIBPmanual
●al.areaLoosecufforcuffhose.●Checkthecuffandcuffhose.