Messages
Message
Location
ExplanationWhattodo
Callservice:Errorx
wherex=0-99
param.
0=RAMfailure
1=ROMchecksumfailure
2=+15Vfailure
3=-15Vfailure
6=ADCerror
7=Watchdogtimetooshort
8=Watchdogtimetoolong
9=Watchdogactivated
10=EEPROMchecksumerror
11=Autozerorangeexceeded
12=Communicationwatchdog
activated
13=Notinuse
14=Tooearlyautocyclestart
Contactauthorizedservice
personnel.
CheckNIBPal.area
Systolicand/ordiastolicresults
missing.
Checkthepatientstatus.
CheckNIBPcuffandhoses.
Repeatthemeasurement.
Controlmeasuremental.areaPressurealarmlimitexceeded.Allowmeasurementto
complete.
Checkthepatientstatus.
Cuffocclusion
param.
Occlusionduringmeasurement
oroverpressuredcuff.
Checkthecuff.
Cuffloose
param.
Loosecufforcuffhose.Checkthecuffandcuffhose.
Cuffoverpressure
param.
NIBPcuffissqueezedduring
measurement.
CheckNIBPcuffandhoses.
Repeatthemeasurement.
Incorrectin.limits
param.
Adultorchildcuffisused,butthe
selectedinfantmoderestricts
theinationpressuretoolowto
beabletomeasuretheblood
pressure.
AdjustNIBPsetup.
Longmeas.time
param.
Themeasurementtimeislong.
Thetriggeringvaluesvary
accordingtothemoduleand
inationlimitsinuse:
PSM:>2minforadult/child,
75sto80sforinfant
PDM:>2minforadult/child,
85sforinfant
TRAM:<3minforadult/child,
90sforinfant
Checkthepatientstatus.
Checkthecuffandhose
connections.
Restartthemeasurement.
Iftheproblempersists,
contactauthorizedservice
personnel.
NIBPairleakage/
NIBPmanual
al.areaLoosecufforcuffhose.Checkthecuffandcuffhose.
2062971-001CARESCAPEModularMonitors473