Invasivepressures
Invasivepressureanalogoutput
InvasivepressuremoduleAnalogoutputsignal
E-modulesP1channel
PDM
Setstherstavailablearterialchannel(Art,
ABP,Fem,UAC)forinvasivepressureanalog
output.
Iftherearenoarterialchannelsavailable,
therstzeroedchannelwillbeused.
Iftherearenozeroedchannelsavailable,
aatlinewillbetheoutputoftheIP
Analogoutputchannel.
TRAM
Setsrst:Art,ABP,Fem,thenUACbasedon
thechannelnumber.
Iftherearenoarterialchannelsavailable,
P1willbeused.
Single-parameterTram-RacmoduleSetsrst:Art,ABP,Fem,thenUACbasedon
thechannelnumber.
Iftherearenoarterialchannelsavailable,
P1willbeused.
SetsP2asthesecondanalogoutputchannel.
SetsP3asthethirdanalogoutputchannel.
SetsP4asthefourthanalogoutputchannel.
Aboutzeroingtheinvasivepressuretransducers
Priortomonitoring,zerotransducersatthepatient’sphlebostaticaxis.Zeroingthe
pressuretransducersisveryimportantforaccuratepressuremeasurements.To
avoidinaccuratemeasurements,youmustzerothepressuretransducers:
Beforemeasuringinvasivepressures.
Beforeinitiatingtreatmentchangesreliantuponpressuresdata.
Whenusinganewtransducerortubing.
Afterreconnectingthetransducercabletotheacquisitiondevice.
Wheneverthepatient’spositionischanged.
Wheneverthepressurereadingisquestionable.
PressurescanbezeroedindividuallybyselectingZeroonthepressuremenuor
bypressingthezerokeyonthemodules.YoucanzeroallpressuresexceptICPby
selectingZeroAllPressuresonthemainmenu.
YoucanzeroallactivetransducersontheE-modulesbypressingeachZeroP1to
P8keys.
YoucanzeroallactivetransducersonthePDMmodulebypressingthezeroall
key.
NOTE
E-modulesrecordatimestampofthelastsuccessfulzeroing
foreachinvasivebloodpressurechannel.
2062971-001CARESCAPEModularMonitors239