Viewingothermonitoredpatients
5.Youcanselecttoseealistofallpatientbedsinthecareunitoralistofremote
patientbedsconguredforalarmnotication.SelectanoptionfromtheShowlist:
●Toshowalistofalltheremotebedsinthecareunit,selectAllPatients.
●Toshowthelistofremotepatientsconguredforalarmnotication,select
NoticationPatientsOnly.
6.Selectapatientbedfromthedisplayedlist.
7.SelectView.
8.Youcanstopviewingtheselectedpatientbedandclosethebed-to-bedwindow
byselectingCloseView.
NOTE
Selectingthehomekeywillnotcloseanyopenbed-to-bed
viewofanalarmingornon-alarmingpatientbed.
Audiopauseforaremotepatientbedalarms
NOTE
ThisfeatureisavailablewiththeCARESCAPENetworkonly.
YoucanpauseactivealarmsfortheremotebedviewedbyselectingRemotefrom
thebed-to-bedwindow.Iftheselectionisnotselectable,thisoptionhasnotbeen
enabledduringconguration.ItisenabledintheCareUnitSettingsanditispassword
protected.Inaddition,theremotemonitor’ssettingsmaypreventremotesilencing
fromtakingplace.
Manualprintingofremotebedwaveforms
NOTE
CARESCAPENetworkonly.
WaveformdataofaremotemonitorcanbemanuallyprintedbyselectingPrintfrom
thebed-to-bedwindow.WiththeB850,waveformsareprintedusingthelocalPRN-50
recorderifavailable,otherwisetheprinterisdeterminedbytheremotemonitor’s
printconguration.Thewaveformsthatappearontheprintoutaredeterminedby
theremotemonitor’sprintconguration.
446CARESCAPEModularMonitors2062971-001