Entropy
Entropyisnotaparameterformonitoringneuromuscularblockade.Though
REmaygiveanindicationofthepatient’sreactiontoexternalstimuli,suchas
intubationandskinincision,thelevelofneuromuscularblockadeshouldbe
assessedwithNMT,whichisanactiveassessmentoftheeffectsofneuromuscular
blockadeagentsontheneuromuscularjunction.
Neuromuscularblockingagents(NMBA)administeredinsurgicallyappropriate
dosesarenotknowntoaffecttheEEG,butareknowntohaveaneffectonthe
EMG.REvaluesmaydropinresponsetoNMBAadministration,duetoparalysis
offacialmuscles.
TheEntropysensorsandcablearedesignedtobedebrillation-proof.
Entropytroubleshooting
ProblemSolution
EntropyvaluesseemunstableCheckthatthesensorisnotdriedout.
Checkthesensorattachmentandplacement.
Checkthepatientstatus.
EntropyEEGsignalisnoisyRemovedisturbingequipmentfromtheproximity
oftheEntropymoduleorsensor.
Checkthesensor’scontactwithskin.
Checkelectrodes.
EntropyEEGsignalispoorCheckthesensor’scontactwithskin.
Checkelectrodes.
EntropyEEGwaveformandnumbersdonot
correspond
CheckrawEEGasimpedancecheckmaycausea
temporaryincreaseinthenumericvalues.
Checkthepatient’soverallstatus.
Entropyreadingsseeminconsistentwiththepatient
status
CheckrawEEGforQRSorotherartifact.
Checkelectrodeplacement.
Thereisasuddendropinvalues.Asuddendropinvaluescanbecausedbythe
following:
Bolusadministrationofintravenousanesthetics.
Increaseininhalationanesthesialevel.
Administrationofothermedicationaffecting
EEG/FEMG.
Thereisanunexpectedincreaseinvalues.Anunexpectedincreaseinvaluescanbecausedby
thefollowing:
Changeininfusionpumpsettingsforintravenous
anesthetics.
Changeinvaporizersettings,orfresh-gasow
rate.
Volumeloadingwithinfuseduids.
Faultintheoperationofanestheticdelivery
systems.
Impedancecheck.
358CARESCAPEModularMonitors2062971-001