PatientSpirometry
x1=Vol
x2=Time
y=Flow
1.Startofinspiration
2.Inspiration
3.Pause
4.Expiration
NormalFlow-Volloop
Whenusingvolume-controlled,constant-owventilationmode,normalFlow-Volloop
reectscorrectfunctioningoftheventilatorandnoairowlimitationsofthepatient.
x=Vol
y=Flow
PatientSpirometrypracticalities
CorrectplacementofthePatientSpirometryaccessoriesisessentialforaccurate
andtrouble-freeoperationofthemeasurement.
TheD-lite(+)/Pedi-lite(+)sensorisplacedclosetothepatient’sairway;rememberto
useD-lite(+)forpatientswhosetidalvolumeis15mlto2000ml,andPedi-lite(+)for
patientswhosetidalvolumeis15mlto300ml.Remembertoselectthecorrect
sensortype.
HumiditycondensinginthePatientSpirometryportsoftheD-lite(+)/Pedi-lite(+)
sensormayincreasethemeasuredvolumesduringlongerlowowanesthesia.
Positionthesensorinsuchawaythatgravitycanremovethecondensedwater
fromthePatientSpirometrytubes.
336CARESCAPEModularMonitors2062971-001