Invasivepressures
3.Setthealarmlimits.
Invasivepressuresalarmpriorities
YoucanselectprioritiesfortheArthigh/low,ABPhigh/low,Femhigh/low,FemV
high/low,CVPhigh/low,PAhigh/low,RAPhigh/low,RVPhigh/low,LAPhigh/low,
ICPhigh/low,CPPhigh/low,andP1high/lowtoP8high/lowalarmsthroughAlarm
Setup>AlarmPriorities>InvasivePressures.ThechoicesareLow,Medium,High,
andEscalating.
Systolicpressurevariationandpulsepressure
variation
Systolicpressurevariation(SPV)andpulsepressurevariation(PPV)canprovideuseful
informationforexamplewhenassessingtheeffectsofuidtherapyonthecardiac
outputofapatient.AparameterwindowwithbothSPVandPPVvaluesappearson
thedisplayprovidedthatSPVhasbeenselectedtothescreenandthearterialsite
selectedastheSPVsourceisactive.
TheSPVandPPVmeasurementisautomatic,andSPVcanalsobetakenmanually.
WiththeNICUsoftwarepackage,noautomaticSPVorPPVareavailable,onlythe
manualSPVcanbeused.
NOTE
TheSPVandPPVmeasurementsarereliableformechanically
ventilatedpatientswithnoarrhythmias,andwhenthearterial
siteselectedastheSPVsourceisprovidingreliablereadings.
ChangingtheSPVsource
1.Selecttheinvasivepressureparameterwindow.
2.SelectArt,ABP,FemorUAC.
3.SelecttheSetuptab.
4.SelectSPVSource>Art,ABP,Fem,orUAC.Youcanalsoturnoffthemeasurement
byselectingOff(default).
MeasuringSPVmanually
TheSPVcanalsobemeasuredmanually.InadditiontoECG1andtheselectedSPV
source,oneofthefollowingparametersisdisplayedinthisorder:Paw,CO
2
,Resp.You
cansettheSPVcursorstodenethedifferencebetweentheminimumandmaximum
systolicpeakpressures.
1.SelecttheSPVandPPVparameterwindow.
2.SelectFreezeWaveforms.
3.Adjustthecursorswiththearrowselectors.
NOTE
YoumustalwaysadjustthecursorsintheNICUsoftware
package.Themonitordoesnotsuggestanycursor
positionsasthereisnoautomaticSPVmeasurementin
thissoftwarepackage.
2062971-001CARESCAPEModularMonitors245