Trends
Changingthegraphictrendscales
1.SelectTrends.
2.SelectGraphicfromtheViewlist.
3.SelectTrendScales.
4.SelecttheGeneral,IP/NIBP,CardiacOutput,orTemptab.
5.Setthetrendscalesforrequiredparameters.
Printingcurrentlyviewedgraphictrends
1.SelectTrends.
2.SelectGraphicfromtheViewlist.
3.SelectPrintPage.
4.YoucanstopprintingbyselectingCancelPrinting.
Printingallgraphictrenddata
1.SelectMonitorSetup>Printing.
2.SelectReports>Trends.
3.SelectPrint.
4.YoucanstopprintingbyselectingCancelPrinting.
Graphictrendresolutionandthehigh-resolutionlicense
Thegraphictrendresolutiondependsonthetimescaleofthetrend.Graphictrends
areupdatedonceaminutewhenthetimescaleis1hourormore.Forthe20minute
scale,theupdaterateis10seconds.
High-resolutiontrendsisalicensedgraphictrendoptionthatprovidesanincreased
resolutionof2secondsforthe2,4,and20minutescales.Thehighresolutiongraphic
trendisupdatedonuserrequest.ThistrendalsocontainscompressedCO
2
and
impedancerespirationwaveforms(10Hz)aswellasbeat-to-beatECGheartrate
andmeanarterialpressureMAP(6Hz).
High-resolutiontrenddataisnotsavedoverpowerdownsituations.Thismeansthat
datainthe2,4,and20minutetimescalesiserased.
High-resolutiontrenddataisneithersentnorloadedto/fromthenetwork(central
station)oracquisitionmodules(PDM,TRAM).
Numerictrends
Viewingnumerictrends
Numerictrendscontainninepageswith24or72hoursoftrenddatadependingon
thetrendlicense.Thetopoftheviewshowsthehighestpriorityrealtimewaveform.
Thelowestrow,Mark,showssnapshoteventnumbers.Ifmorethanonesnapshothas
beencreatedinaone-minuteperiod,onlythelastsnapshoteventnumberisshown.
YoucannotcongurethelayoutoftheNumerictrendview.
1.SelectTrends.
2062971-001CARESCAPEModularMonitors419