Alarms
Powerloss
duration
Alarmsettingstatusafterapowerloss
Upto15minutes
Thealarmsettingsthatareineffectbeforethepowerlossarerestored
automatically.
Greaterthan15
minutes
Thealarmsettingsrevertbacktotheuserdefaultsettings(start-up
mode).Youmustrecongureanypatient-specicalarmsettings.
AlarmdatastoredinClinicallogs
AccesstotheClinicallogsisaservice-levelfunctionanditispasswordprotected.
Themonitorstoresarecordofpatient-relatedlocalandremotealarmsand
informationmessagesaswellasanyadjustmentstothealarmlimitsintheClinical
logs.
Storedalarmdataduringapowercycleorpower
loss
Ifthemonitorgoesthroughapowercycleoralossofpower,thestoredalarmdata
isnotaffected.ThealarmdataremainsstoredintheClinicalloguntilthemonitor
automaticallyclearstheoldeststoreddatatoallownewdatatobestored.
2062971-001CARESCAPEModularMonitors145