Cleaningandcare
3.Forgeneralcleaningofcablesandleadwires,wipeusingalightlymoistenedcloth
withamildsoapandwatersolution.
4.Fordisinfectingthecablesandleadwires,wipeexteriorwithasoftlint-freecloth,
usingachemicaldisinfectant.Refertothecables’andleadwires’owninstructions
forusefordetailedinformationregardingallowedsubstances.
Wringexcessdisinfectantfromwipebeforeusing.
Anycontactofdisinfectantsolutionswithmetalpartsmaycausecorrosion.
Donotimmerseeitherendofacableorleadwireconnector.Immersingorsoaking
theconnectorendsmaycorrodemetalcontactendsandaffectsignalquality.
5.Wipeoffcleaningsolutionswithaclean,lightlymoistenedcloth.
6.Drythoroughlywithadry,lint-freeclothandletairdryforatleast30minutes.
Dryingtimesmayvarybasedontheenvironmentalconditions.Donotapplyheat.
ReusableD-liteandPedi-litesensorcleaninginstructions
ReusableD-liteandPedi-litesensorscanbewashed,disinfected,orsteamautoclaved.
Makesurethatthesensorisdryandtheconnectorsarenotdamaged.Atight
connectionisessentialforcorrectmeasurement.
D-fend(+),D-fendPro(+),andMiniD-fendwatertrapcare
instructions
Emptythecontainerwheneveritismorethanhalffull.
Inanesthesia:ReplacetheD-fend,D-fendPro,orMiniD-fendwatertrapeverytwo
monthsorwhenReplaceWaterTrapappears.
Incriticalcare:ReplacetheD-fend+,D-fendPro+,orMiniD-fendwatertrapevery
24hours,foreachnewpatient,orwhenthemessageReplaceWaterTrapappears.
Whentakinganewwatertrapintouse,markthedateontheappropriatelabelon
thewatertrapcartridge:
Thewatertrapcartridgeisdisposable.Donotwashorreusethecartridge.
CAPNOSTATsensorandadaptercleaninginstructions
B850only.
Cleanthesensorsurfaceswithadampclothandallowtodry.Donotimmersethe
sensororattempttosterilizethesensor.Beforereusing,makesurethesensor
windowsarecleananddry.
Cleantheadapterbyrinsinginawarmsoapysolution,followedbysoakingina
liquiddisinfectant,pasteurized,orcoldsterilized(Glutaraldehyde).Rinsethoroughly
withsterilizedwaterandallowtodry.Beforereusingtheadapter,makesurethe
adapterwindowsarecleananddry.
HowtostorePDMandPSM
RemovePDMbatterieswhenthedeviceisnotinuse,evenforshortperiodsoftime.
458CARESCAPEModularMonitors2062971-001