Invasivepressures
SelectingthedisplaymodeforIPwaveforms
Youcanselecttheinvasivepressurewaveformstobeshownasindividualwaveforms,
orinacombinedview.
1.SelectMonitorSetup>ScreenSetup.
2.B850andB450withtheDoubleVideolicense:SelectScreen1orScreen2tab.
3.SelectUpperParameterArea.
4.SelectanoptionfromtheInvasivePressureWaveformslist:
Toviewindividualwaveforms,selectIndividual.
Tocombinethecurrentlydisplayedadjacentwaveforms(2to4),select
Combined.Thenewwaveformeldwillusethecombinedheightoftheoriginal
elds.
Tocombineuptofourwaveformsinoneeld,select4invP.Thenewwaveform
eldwillusetheheightoftwoupperparameterwindows.
Usingtheinvasivepressuremeasurement
Invasivepressuremeasurementmapping
Invasivepressuremeasurementsaremappedtooneofeight(B850),seven(B650),or
six(B450)invasivepressurechannelsasfollows:
Pressurechannel
Pressuremeasurementsource
P1
PDM(Masimo),PDM(Nellcor),TRAM451,TRAM451N5,TRAM
451M,E-PSMP.
P2
PDM(Masimo),PDM(Nellcor),TRAM451,TRAM451N5,TRAM
451M,E-PSMP.
P3
PDM(Masimo),PDM(Nellcor),TRAM451,TRAM451N5,TRAM
451M,.
P4
PDM(Masimo),PDM(Nellcor),TRAM451,TRAM451N5,TRAM
451M,.
P5
DualBP,BP/CO,orBP/DualTempTram-RacmoduleinTram-Rac
slot3,orE-PP.
P6
DualBPorBPTram-RacmoduleinTram-Racslot3,E-PP.
P7
DualBP,BP/CO,orBP/DualTempTram-RacmoduleinTram-Rac
slot4,E-P,orE-PT.
P8
DualBPorBPTram-RacmoduleinTram-Racslot4,E-COP,or
E-COPSv.
238CARESCAPEModularMonitors2062971-001