Cardiacoutput
catheterswhichshowupinthelistcanbeaddedordeletedthroughCareUnit
Settings>Parameters>Catheters.Thesesettingsarepasswordprotected.
Formoreinformation,seetheCARESCAPEModularMonitorsSupplemental
InformationManual.
SelectingaC.O.catheterfromthelist
1.Selectthecardiacoutputparameterwindow.
2.SelecttheSetuptab.
3.SelectacatheternamefromtheManufacturerlist.
4.SelectacathetermodelfromtheModellist.
Enteringauser-denedC.O.catheter
Alluser-denedcathetersettingsareerasedwhenthemonitorisdischarged.
1.Selectthecardiacoutputparameterwindow.
2.SelecttheSetuptab.
3.SelectUserDenedfromtheManufacturerlist.
4.SettheInjectateVolumetomatchthevaluelistedonthecatheterpackaging.
5.SettheComputationConstanttomatchthevaluelistedonthecatheter
packaging:
●IceCold:temperaturebelow+6°C.
●RoomTemp:temperateabove+18°C.
●Anyotherinjectatetemperaturevaluebetween+6°Cand+18°C.
SelectingtheC.O.injectateprobetype
NOTE
PDMandTRAMonly.E-modulesdetectthetypeofinjectate
probeautomatically.
1.Selectthecardiacoutputparameterwindow.
2.SelecttheSetuptab.
3.SelectthecorrectProbeType:BathorIn-Line.
SettingaC.O.rightventricularejectionfraction(REF)
measurement
NOTE
E-COPandE-COPSvmodulesandcathetersthatsupportright
ventricularejectionfractionmeasurementonly.PDMand
TRAMdonotprovideaREFmeasurement.
AvalidheartrateisrequiredtotakeaREFmeasurement.ECGfromatelemetry
transmittercannotbeused.
1.Selectthecardiacoutputparameterwindow.
2.SelecttheSetuptab.
3.SelectthecheckboxforREFMeasurement.
2062971-001CARESCAPEModularMonitors267