GE CARESCAPE - Selecting the C.O. injectate probe type; Setting a C.O. right ventricular ejection fraction (REF) measure

GE CARESCAPE
548 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cardiacoutput
catheterswhichshowupinthelistcanbeaddedordeletedthroughCareUnit
Settings>Parameters>Catheters.Thesesettingsarepasswordprotected.
Formoreinformation,seetheCARESCAPEModularMonitorsSupplemental
InformationManual.
SelectingaC.O.catheterfromthelist
1.Selectthecardiacoutputparameterwindow.
2.SelecttheSetuptab.
3.SelectacatheternamefromtheManufacturerlist.
4.SelectacathetermodelfromtheModellist.
Enteringauser-denedC.O.catheter
Alluser-denedcathetersettingsareerasedwhenthemonitorisdischarged.
1.Selectthecardiacoutputparameterwindow.
2.SelecttheSetuptab.
3.SelectUserDenedfromtheManufacturerlist.
4.SettheInjectateVolumetomatchthevaluelistedonthecatheterpackaging.
5.SettheComputationConstanttomatchthevaluelistedonthecatheter
packaging:
IceCold:temperaturebelow+6°C.
RoomTemp:temperateabove+18°C.
Anyotherinjectatetemperaturevaluebetween+6°Cand+18°C.
SelectingtheC.O.injectateprobetype
NOTE
PDMandTRAMonly.E-modulesdetectthetypeofinjectate
probeautomatically.
1.Selectthecardiacoutputparameterwindow.
2.SelecttheSetuptab.
3.SelectthecorrectProbeType:BathorIn-Line.
SettingaC.O.rightventricularejectionfraction(REF)
measurement
NOTE
E-COPandE-COPSvmodulesandcathetersthatsupportright
ventricularejectionfractionmeasurementonly.PDMand
TRAMdonotprovideaREFmeasurement.
AvalidheartrateisrequiredtotakeaREFmeasurement.ECGfromatelemetry
transmittercannotbeused.
1.Selectthecardiacoutputparameterwindow.
2.SelecttheSetuptab.
3.SelectthecheckboxforREFMeasurement.
2062971-001CARESCAPEModularMonitors267