Messages
MessagesrelatedtoECGmeasurement
Forinformationregardingalarmprioritiesandescalationtimes,seetheCARESCAPEModular
MonitorsSupplementalInformationManual.
Makesureyouarefamiliarwiththegenericlayoutofthescreen.Thiswillhelpyouidentify
whereonscreenthefollowingmessagesappear.Themessagelocationisindicatedwiththe
followingabbreviations:
al.area=alarmarea
param.=parameterwindow
report=reportview
wavef.=waveformarea
Message
Location
ExplanationWhattodo
12diagnosticleads
needed
report
SomeoftheECGleadsare
disconnected.
CheckthatallECGleadsare
connectedtothepatient.
ACI-TIPIon-ageless
than16
report
ThesettingisONbutthepatient
isunder16yearsold.
SettheACI-TIPItooffto
create12SLreports.
Adjustthepatientageonthe
monitorifitisnotcorrect.
ACI-TIPIon-chest
orleftarmpainnot
entered
report
ThesettingisONbutchestor
leftarmpaininformationhas
notbeenentered.
SettheACI-TIPItooff.
Enterchestorleftarmpain
information.
AFibal.area,
wavef.
Physiologicalalarm.Checkthepatientstatus.
Accel.Ventric.al.area,
wavef.
Physiologicalalarm.Checkthepatientstatus.
Arrhoff
param.
Thearrhythmiadetectionlevel
issettoOff.
Ifarrhythmiadetectionis
needed,setthedetection
leveltoFullorLethal.
Arrhythmiapaused
Arrhpaused
al.area
wavef.
ECGchannelshavenotbeen
availableforanalysisforthe
last20secondsortheinternal
HRcalculationhasnotbeen
updatedforthelast30seconds
duetoexcessiveartifact.
Checkthepatientstatus.
Checkelectrodeplacement.
Preparethepatient’sskinat
electrodesites.
Changeormoveelectrodes.