Messages
Message
Location
ExplanationWhattodo
Artifactwavef.Muscleartifactorhighorlow
frequencynoise.
Checkelectrodecontact.
Checkleadplacement.
Performskinpreparation.
Reposition/replace
electrodes.
Requestthepatienttoremain
still.
Asystoleal.area,
param.,
wavef.
Physiologicalalarm.Checkthepatientstatus.
Bigeminy
al.area,
wavef.
Physiologicalalarm.Checkthepatientstatus.
Bradyal.area,
wavef.
Physiologicalalarm.Checkthepatientstatus.
Cableoff
report
ECGcableisoff.Connectthecable.
Changetelemetry
battery
al.areaThetelemetrytransmitter’s
batteryhaslittlechargeleft.
Changethebattery.
DuplicateTTXal.areaMultiplepatientshavebeen
admittedusingthesameTTX
number.
Contactauthorizedservice
personnel.
ECGerror
al.areaThemodulehasa
communicationerror.
Checkthemodule.
Replacethemoduleif
necessary.
ECGmeasurements
removed
al.areaAcquisitionmodulehaslostECG
communication.
Removethemoduleandthen
reconnectit.
Replacethemodule.
Iftheproblempersists,
contactauthorizedservice
personnel.
Failureconnectingto
transmitter
al.areaConnectiontothetelemetry
transmittercouldnotbe
established.
Retryconnecting.
Iftheproblempersists,
contactauthorizedservice
personnel.
FrequentPVCs
al.areaPhysiologicalalarm.Checkthepatientstatus.
FrequentSVCs
al.area,
wavef.
Physiologicalalarm.Checkthepatientstatus.
Genderisnotdened
report
ACI-TIPIhasbeenselectedon,
butthepatient’sgenderhasnot
beenentered.
Enterthepatient’sgender.
HR(ECG)high/
HR(ECG)low
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
464CARESCAPEModularMonitors2062971-001