Location
ExplanationWhattodo
HRhigh/HRlow
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
Irregularal.area,
wavef.
Physiologicalalarm.Checkthepatientstatus.
LA/Lleadoff
LL/Fleadoff
RA/Rleadoff
RL/Nleadoff
wavef.Anelectrodeisdisconnected.Checktheelectrodes.
Leadchangedwavef.Themonitorautomatically
switchestheECG1waveform
selectiontoameasurableECG
Lead(I,II,III,aVR,aVL,aVF,
V1,V2,V3,V4,V5orV6)ifthe
currentECG1waveformisnot
measurable.
NotethattheECGwaveform
changesaccordingtothe
leaditismeasuredfrom.
Checkthelead.
Leadsoffal.area,
wavef.
Oneormoreoftheconnected
electrodesisdisconnectedand
arrhythmiadetectionisnot
possible.
ECG:Checktheconnections.
Learning
wavef.STalgorithmisinlearningphase,
messageshowne.g.whenECG
measurementisstarted.
Noactionrequired.
Missingbeatal.area,
wavef.
Physiologicalalarm.Checkthepatientstatus.
MultifocalPVCsal.area,
wavef.
Physiologicalalarm.Checkthepatientstatus.
No10or6leadcable
report
No6-or10-leadECGcableis
attached.
Connecta6-or10-leadcable.
No12RLlicense
report
No12RL12leadECGlicense.
ContactyourGE
representativetopurchase
thisfeature.
No12SLlicense
report
No12SLECGwithACI-TIPI
license.
ContactyourGE
representativetopurchase
thisfeature.
Notelemetryal.areaTheservercannotcommunicate
withthetelemetrytransmitter.
Makesurethetransmitteris
notoutofrange.
Checkthetelemetry
transmitter’sbattery.
Waittoseeiftheproblem
resolves.Iftheproblem
persists,contactauthorized
servicepersonnel.