9
Impedancerespiration
Impedancerespirationcompatibilitylimitations
TRAMandTram-RacmoduleswiththeB850only.
Fordetailedinformationregardingmodule,monitor,andaccessorycompatibility,see
theCARESCAPEModularMonitorsSupplementalInformationManual.
Respirationsafetyprecautions
Respirationwarnings
WARNING
Makesurethattheleadwiresetclipsorsnapsdonottouch
anyelectricallyconductivematerialincludingearth.
WARNING
Theimpedancerespirationmeasurementisinherently
verysensitiveasitmeasuresverysmallphysiological
signals(changesofimpedanceofthepatient’schest
area).Electromagneticinterferencemaycauseerroneous
measurementsatvariousfrequencies,forexample
interferencewiththesignal/waveform,leadingtorespiration
ratereadingsinconsistentwiththepatient'struerespiration
rate.Ifyounoticethis,useanotherformofrespiration
monitoring,forinstanceend-tidalCO
2
.
WARNING
Whenusinganelectrosurgeryunit,notethatthe
measurementcablesdonotincorporatemeanstoprotect
againstburnsincaseofadefectiveESUreturnelectrode.To
avoidburnsatthemonitormeasurementsites,ensurethe
following:
●PropercontactoftheESUreturnelectrodetothepatient.
●ESUreturnelectrodeneartheoperatingarea.
●Measurementelectrodes,leadwiresandprobesfarfrom
thesurgicalsiteandtheESUreturnelectrode.
WARNING
—APNEAEVENTS—Themonitormaynotdetectallepisodes
ofinadequatebreathing,nordoesitdistinguishbetween
central,obstructive,andmixedapneaevents.
2062971-001CARESCAPEModularMonitors191