EasyManua.ls Logo

GE CARESCAPE - Alarm priority levels; Alarm priority escalation; Physiological alarms’ activation criteria

GE CARESCAPE
548 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Alarms
Technicalalarmconditionsaretriggeredbyanelectrical,mechanical,orother
failureoftheequipment,orbyfailureofasensororcomponent.Technicalalarm
conditionsmayalsobecausedwhenanalgorithmcannotclassifyorinterpretthe
availabledata.Thevisualmanifestationofatechnicalalarmisactiveaslongas
thereasonforthatalarmexists.
Certaintechnicalalarmscanbedeactivatedwiththepauseaudiokey.The
AlarmDeactivationsettingisconguredintheCareUnitSettingsandit
ispasswordprotected.Formoreinformation,seetheCARESCAPEModular
MonitorsSupplementalInformationManual.
Alarmprioritylevels
Physiologicalandtechnicalalarmsarecategorizedbyprioritylevel:
Highpriorityalarmsrequireanimmediateresponse.
Mediumpriorityalarmsrequireapromptresponse.
Lowpriorityalarmsrequireyoutobeawareofthiscondition.
Informationalprioritymessagesprovideinformationyoushouldknow.
NOTE
Informationalmessagesarenotsenttothenetwork,and
theyareneverlatched.
Alarmpriorityescalation
Anescalatingalarmstartsatadesignatedprioritylevel(lowormedium)andwill
escalatetothenexthigherprioritylevelofalarm(afterasetnumberofseconds)ifthe
alarmconditionhasnotbeenresolved.Itisimportanttonotethattheseescalateup
tothenextlevelbutwillnotresetuntiltheconditionhasbeenresolved.
NOTE
Alarmpriorityescalationaffectsthecurrentlyongoingalarm
condition,notanyfuturealarmsofthesametype.Anynew
alarmswillalarmattheirdesignatedprioritylevel,notatthe
escalatedlevel.
Formoreinformation,seetheCARESCAPEModularMonitorsSupplemental
InformationManual.
Physiologicalalarms’activationcriteria
Physiologicalalarmshaveindividualactivationcriteriaasshowninthetable.Alarm
annunciationdoesnotdependoncaseactivity.
Parameter
Alarmactivationcriteria
ECG
Activemeasurementfor30seconds.
ImpedancerespirationActivemeasurementfor30seconds.
SpO
2
Activemeasurementfor60seconds.
NIBP
Manual,Auto,orStatmodestarted.
IP
Activemeasurementfor30seconds.
Temperature
Activeassoonasmeasurementreadingsareavailable.
C.O.
Activeassoonascontinuousmeasurementreadingsareavailablefromthe
UnityNetworkInterfaceDevice(ID).
130CARESCAPEModularMonitors2062971-001

Table of Contents