EEGandauditoryevokedpotentials
ECG:Cardiacelectricalactivitycanaffect
theEEGwaveform.
EMG:Activationoffrontalmusclescan
causehighfrequencyinterferenceinthe
EEG.
EEGtroubleshooting
ProblemSolution
WhatcanIdoifthesignallooksnoisyandthereisa
messageindicatingartifact?
Electrodesarepoorlyconnected,orelectrical
interferenceiscouplingtoelectrodecables.
Checkthattheelectrodesareproperlyconnected
andnotdriedout.
Checktheelectrodes’contactwithskin.
Performelectrodeimpedancecheck.
Calmthepatientsincefrontalmuscleactivitycan
causeartifact
Removesourcesofexternalelectricalnoise(for
example,somelamps)fromthevicinityofthe
patient’shead.
ECGmaycauseartifact;changeelectrode
positioning.
Checkthattherearenottwoidentical
measurementmodulesinthesystem.
WhyareallEEGwaveformsnotdrawnevenif
electrodesandcablesareOK?
Thenumberofchannelsinthemontageissmaller
thanthenumberofchannelsconnectedtothe
patient.
386CARESCAPEModularMonitors2062971-001