ECG
3.SelectareferenceQRSfromtheEraseReferencelist.
IfyoudeletethereferenceQRScurrentlydisplayed,thenthenext,newerreference
QRSisdisplayedasthereferenceQRS.
PrintingarealtimeQRS/STreport
Thisselectionisnotavailablewithcombinationmonitoring.
TheQRS/STreportdisplaysthecurrentSTleadsandthetrendsat10minuteintervals.
1.SelecttheSTparameterwindow.
2.SelectRealtimeView.
3.SelectPrintQRS/ST.
4.Tostopprinting,selectCancelPrinting.
ViewingQRSandSTinasplitscreen
Inthesplitscreenwindow,youcanviewthereferenceQRScomplex,currentQRS
complexes,andSTtrends.
1.SelectMonitorSetupfromthemonitor’smainmenu.
2.SelectScreenSetup.
3.SelectSplitScreen
4.SelectSTtoviewanST/QRSsplitscreen.
SelectingtheSTtimescale
Thisselectionisnotavailablewithcombinationmonitoring.
ThissettingalsodeterminesthelengthoftheSTtrendreport,andyoucanselectit
fromtheRealtimeViewortheTrendView.
1.SelecttheSTparameterwindow.
2.SelectRealtimeVieworTrendView.
3.SelectavaluefromtheTrendScaleslist.
STtrenddisplay
Thisfeatureisnotavailablewithcombinationmonitoring.
EachQRSwindowdisplaysthecurrentQRScomplexandSTvalue.TheQRScomplexes
andcurrentmeasurementpointlinesareupdatedatleastevery40seconds.Each
QRSwindowhasacorrespondingtrendwindowdisplayingtheSTtrendalongwith
thetrendscaleandcurrenttimeinterval.TheprimaryHRtrendwindowisdisplayed
belowthelastSTtrendwindow.TheSTtrendsfortheavailableleadsareupdated
every10seconds.