GE CARESCAPE - Selecting Smart BP; Compensating for intra-aortic balloon pump (IABP) waveform irreg; Selecting the invasive pressure response time

GE CARESCAPE
548 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Invasivepressures
2.SelecttheSetuptab.
3.SelectShowCPP.
SelectingSmartBP
NOTE
PDM,TRAMandTram-Racmodulesonly.Art,ABP,andFem
invasivepressurechannelsonly.
SmartBPisanalgorithmthattemporarilydeactivatesthearterialandfemoralalarms
whenitdetectsthezeroingofatransducer,fastushingofthesystem,orblood
draws.ThemessageArtifactdisplaysduringthealarmdeactivation.Whenpulsatile
pressurereturnsand15-20beatsaredetected,numericsaredisplayedandalarms
arereactivated.
1.Selecttheinvasivepressureparameterwindow.
2.SelecttheAdvancedtab.
3.SelectSmartBP.
Compensatingforintra-aorticballoonpump(IABP)
waveformirregularities
WARNING
INCORRECTPULSERATEBesuretoturnofftheIABP
settingwhenthecardiacassistdeviceisnolongerused.
Failuretodosocouldresultinincorrectpulseratereadings.
CAUTION
PATIENTHAZARDIfyouchoosetotriggertheballoon
pumpfromthemonitor,contacttheballoonpump
manufacturerdirectlyforinterfacerequirements,asthey
varyamongmanufacturers.Sometriggermodesoncertain
balloonpumpdevicesmaynotbecompatiblewiththeGE
analogoutputsignal,andusemaycontributetopatientinjury
orsub-optimalpumpingresults.
NOTE
PDM,TRAM,andTram-Racmodulesonly.Art,ABP,andFem
invasivepressurechannelsonly.
1.Selecttheinvasivepressureparameterwindow.
2.SelecttheAdvancedtab.
3.SelectIABPOn.
IABPnowdisplaysintheinvasivepressurechannelparameterwindow.
Selectingtheinvasivepressureresponsetime
NOTE
E-modulesonly.
1.Selecttheinvasivepressureparameterwindow.
2.SelecttheSetuptab.
3.SelectNormalorB-to-B(beat-to-beat)fromtheResponselist.
2062971-001CARESCAPEModularMonitors243