Invasivepressures
2.SelecttheSetuptab.
3.SelectShowCPP.
SelectingSmartBP
NOTE
PDM,TRAMandTram-Racmodulesonly.Art,ABP,andFem
invasivepressurechannelsonly.
SmartBPisanalgorithmthattemporarilydeactivatesthearterialandfemoralalarms
whenitdetectsthezeroingofatransducer,fastushingofthesystem,orblood
draws.ThemessageArtifactdisplaysduringthealarmdeactivation.Whenpulsatile
pressurereturnsand15-20beatsaredetected,numericsaredisplayedandalarms
arereactivated.
1.Selecttheinvasivepressureparameterwindow.
2.SelecttheAdvancedtab.
3.SelectSmartBP.
Compensatingforintra-aorticballoonpump(IABP)
waveformirregularities
WARNING
—INCORRECTPULSERATE—BesuretoturnofftheIABP
settingwhenthecardiacassistdeviceisnolongerused.
Failuretodosocouldresultinincorrectpulseratereadings.
CAUTION
—PATIENTHAZARD—Ifyouchoosetotriggertheballoon
pumpfromthemonitor,contacttheballoonpump
manufacturerdirectlyforinterfacerequirements,asthey
varyamongmanufacturers.Sometriggermodesoncertain
balloonpumpdevicesmaynotbecompatiblewiththeGE
analogoutputsignal,andusemaycontributetopatientinjury
orsub-optimalpumpingresults.
NOTE
PDM,TRAM,andTram-Racmodulesonly.Art,ABP,andFem
invasivepressurechannelsonly.
1.Selecttheinvasivepressureparameterwindow.
2.SelecttheAdvancedtab.
3.SelectIABPOn.
IABPnowdisplaysintheinvasivepressurechannelparameterwindow.
Selectingtheinvasivepressureresponsetime
NOTE
E-modulesonly.
1.Selecttheinvasivepressureparameterwindow.
2.SelecttheSetuptab.
3.SelectNormalorB-to-B(beat-to-beat)fromtheResponselist.
2062971-001CARESCAPEModularMonitors243