Neuromusculartransmission
2.SelectStartwith>Currentpatient.
3.SelectContinue.
RestartingtheNMTmeasurementinORafterinduction
WhenyoumovethepatientwiththemoduletotheORandwanttocontinuethe
measurementwiththealreadyfoundanddeterminedcurrentandreferencevalues,
usetherestartfunction.ConnectthemoduletothemonitorandpresstheStop
Continuemodulekey,or:
1.SelecttheNMTparameterwindow>Setup.
2.SelectStartwith>Recallreference.
3.SelectRestart.
StoppingtheNMTmeasurement
PresstheStopContinuemodulekey,or:
1.SelecttheNMTparameterwindow>Setup.
2.SelectStop.
NMTalternativeuses
Localnerveandplexuslocalization
Singlestimulationpulsesmaybehelpfulfordeterminingthecorrectneedletipposition
fromalocalnerveinplexusprocedures.TheNMTmoduledeliverssinglestimulation
pulsesataselectedrateuntilitismanuallystopped.Musclecontractionsfromthe
stimulatedinnervatingmotoricnerve(s)maybeobserved.Notethatyouneedspecic
accessoriesforthismeasurement.
2062971-001CARESCAPEModularMonitors367