Startingandendingmonitoring
Endofstandby
Themonitorendsthestandbyautomaticallywhenanyofthefollowingconditions
occur:
VitalssignsarestillpresentafterPrepareforStandbyhasbeenselectedandthe
audiopausetimeexpires.
Anyvitalsignsaredetectedasactive.
Userinputisreceived:akeyboardkeyispressed,TrimKnobispressedorrotated,
primarymousebuttonispressed,touchscreenispressed.
APDMorTRAMisconnected.
Aboutcontinuingmonitoring
PatientinformationanddataisstoredinthePDMandTRAM.TheConnecting
MeasurementmessageappearsonthemonitorwhenthePDMorTRAMisrst
connected.Vitalsignmonitoringisnotavailableduringthisinitializationtime.At
theconclusionoftheinitialization,thestoredpatientinformationanddatacanbe
transferredtothemonitorfromtheacquisitionmodule.Whilepatientinformationand
dataisbeingtransferred,theLoadingfromPDMorLoadingfromTrammessage
appearsonthemonitor.Thetypeofpatientinformationanddatathatcanbe
transferredincludesthefollowing:
Patientdemographicdata
Patienttrendsandalarmhistories;alarmhistoriesareconvertedtosnapshots
duringthetransfer
Allzeroedinvasivepressuresitelabels
Allinvasivepressuretransducerzerovaluesforchannels1to4
LatestvaluesandtimestampsforC.O.,PCWP,andNIBPmeasurements
NIBPcuffsize
NIBPautocycleon/offinformation
NIBPcycletimeinformationiftheNIBPautocycleison
Theactionstakenbythemonitorandthemenusthatappeardependonwhetherthe
monitorhasanactivepatientcase/admittedpatientornot.
Howtocontinuemonitoringwhenacaseisnot
active/patientisdischarged
WhenyouconnectaPDMorTRAMthatisactivelymeasuringvitalsignsfromECG,
invasivepressuresorSpO
2
,thepatientinformationanddataisautomaticallyloaded
fromthePDMorTRAMtothemonitorandacaseisautomaticallystarted/patientis
admitted.
WhenyouconnectaPDMorTRAMthatisnotactivelymeasuringvitalsignsfrom
ECG,invasivepressureorSpO
2
butcontainspatientinformation,theContinuemenu
appears.TheContinuemenuhastwoinformativeeldsonthetop:thePatientinthe
monitorandthePatientinthePDMorPatientintheTram.Theseeldsshowthetext
Nopatientidenticationdataavailableifthereisnoactivecase/admittedpatient.
Selectoneofthefollowingtocontinuemonitoring:
LoadPDMData/LoadTramData:Thisselectionloadspatientinformationand
datafromthemodule.
2062971-001CARESCAPEModularMonitors123