Cleaningandcare
B650,B450:Checkthechargeofthemonitorbattery.(BatteryisoptionalinB650.)
Monthlychecks
B850andB650only.
Checkthegasexchangecalibration.
Checkeverytwomonths
Changethewatertrap.
Checktheairwaygasescalibrationifthemeasurementisincontinuoususe.
Checkeverysixmonths
Checktheairwaygasescalibrationifthemeasurementisinnormal(notcontinuous)
use.
Onceayearchecks
Checkthecalibrationoftemperature,NIBPandinvasivebloodpressure.
NOTE
Theinvasivebloodpressuretransducersshouldbe
calibratedwheneveratransducererroroccurs.
CheckthePatientSpirometryowcalibration.
Regularcalibrationchecks
Thefollowingparametersrequirecalibrationchecksatregularintervals,inaddition
tothecalibrationperformedwhilemonitoringpatients.
Airwaygases
Therecommendedcalibrationintervalforairwaygasmeasurementsisevery
sixmonthsinnormaluseandeverytwomonthsincontinuoususe,toensure
thatthemeasurementaccuracyremainswithinspecications.Familiarize
yourselfwithinstructionsregardingtheairwaygasesmeasurement,including
thecalibrationinstructions.
PatientSpirometry
Ifthedifferencebetweentheinspiratoryandexpiratoryvolumesispermanent,
aowcalibrationisrequired.Forcalibrationinstructions,seetheModule
FramesandModulesTechnicalManual.
Gasexchange
Therecommendedgasexchangecalibrationintervalisonceamonthtoensure
thatthemeasurementaccuracyremainswithinspecications.Forcalibration
instructions,seetheModuleFramesandModulesTechnicalManual.
Temperature,NIBP,andinvasivepressures
Acalibrationcheckoftemperature,NIBPandinvasivebloodpressuresshould
beperformedatleastonceayeartoensurethatthemeasurementaccuracy
remainswithinspecications.Forcalibrationinstructions,seetheModule
FramesandModulesTechnicalManual.