Cleaningandcare
●B650,B450:Checkthechargeofthemonitorbattery.(BatteryisoptionalinB650.)
Monthlychecks
B850andB650only.
●Checkthegasexchangecalibration.
Checkeverytwomonths
●Changethewatertrap.
●Checktheairwaygasescalibrationifthemeasurementisincontinuoususe.
Checkeverysixmonths
●Checktheairwaygasescalibrationifthemeasurementisinnormal(notcontinuous)
use.
Onceayearchecks
●Checkthecalibrationoftemperature,NIBPandinvasivebloodpressure.
NOTE
Theinvasivebloodpressuretransducersshouldbe
calibratedwheneveratransducererroroccurs.
●CheckthePatientSpirometryowcalibration.
Regularcalibrationchecks
Thefollowingparametersrequirecalibrationchecksatregularintervals,inaddition
tothecalibrationperformedwhilemonitoringpatients.
●Airwaygases
■Therecommendedcalibrationintervalforairwaygasmeasurementsisevery
sixmonthsinnormaluseandeverytwomonthsincontinuoususe,toensure
thatthemeasurementaccuracyremainswithinspecications.Familiarize
yourselfwithinstructionsregardingtheairwaygasesmeasurement,including
thecalibrationinstructions.
●PatientSpirometry
■Ifthedifferencebetweentheinspiratoryandexpiratoryvolumesispermanent,
aowcalibrationisrequired.Forcalibrationinstructions,seetheModule
FramesandModulesTechnicalManual.
●Gasexchange
■Therecommendedgasexchangecalibrationintervalisonceamonthtoensure
thatthemeasurementaccuracyremainswithinspecications.Forcalibration
instructions,seetheModuleFramesandModulesTechnicalManual.
●Temperature,NIBP,andinvasivepressures
■Acalibrationcheckoftemperature,NIBPandinvasivebloodpressuresshould
beperformedatleastonceayeartoensurethatthemeasurementaccuracy
remainswithinspecications.Forcalibrationinstructions,seetheModule
FramesandModulesTechnicalManual.