EasyManua.ls Logo

GE CARESCAPE - Audible and visual alarm signals; Auditory information signals; Audible alarms off behavior

GE CARESCAPE
548 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Alarms
Thelowerthenumber,thequieterthealarmvolume.Notethattheminimumallowed
alarmvolumelevelsaresetintheCareUnitSettings.
Audibleandvisualalarmsignals
Alarmsignalsindicatethatanalarmconditionispresent.Thealarmprioritylevelsare
indicatedbyvisualandaudiblesignals.Thevisualandaudiblealarmsignalsassume
thatthepatientmonitorandtheoperatorarewithinthepatientenvironment(1.5
meters).
WhenphysiologicalalarmmessagesarereceivedfromothermonitorsontheS/5
Network,onlythemostrecentandhighestprioritymessageisshown.Thereareno
ashingnumericsorbackgroundcolor.Nomessagesareshownintheparameter
windowarea.
Thefollowingtablelistsalarmsignalsfordifferentalarmprioritylevels:
Prioritylevel
Signal
HighMedium
Low
Informational
Parameterwindow
physiologicaldatavalues
Blacktextashesinside
aredbox.
Blacktextashes
insideayellowbox.
Blacktextinsidea
cyan(blue)box.
Notapplicable.
AlarmareaWhitetextinsideared
box.
Blacktextinsidea
yellowbox.
Blacktextinsidea
cyan(blue)box.
Blacktextinsidea
greybox.
Waveformeldmessages
TextTextTextText
Audibletonepattern
1
(IEC
60601-1-8)
Repeatspatternof2*5
beeptones
Repeatspatternof
3-beeptones
1-beeptone
None
Audibletonepattern
(legacy)
Repeatspatternof
3-beeptones(crisis)
Repeatspattern
of2-beeptone
(warning)
1-beeptone
(advisory)
None
Alarmlightindicator
2
FlashesredFlashesyellowSolidblueNoeffect
Automaticviewonalarm
audibletonepattern
Usercongurable
3
Usercongurable
3
Usercongurable
3
None
1
TheIECaudibletonepatternisthefactorydefaultsetting.
2
Whentheaudiblealarmsareturnedofforarepaused,thealarmAudiopause/offareaoftheAlarmlightisasolid
bluecolor.
3
TheAVOAaudiblealarmtonepatternmaybeconguredtomatchthelocalmonitor’saudibletonepattern,repeata
patternoftwobeeptones,beeptwiceuponactivationbyanalarmcondition,orbeturnedoff.ForconguringtheAVOA
audiblealarmtonepattern,seetheCARESCAPEModularMonitorsSupplementalInformationManual.
Auditoryinformationsignals
Themonitorperformsaself-diagnosticprocedureatstart-upandgeneratesan
auditorytestsignal.Therearealsootherauditoryinformationsignalsindicatingthe
statusofsomeparametermeasurements.
Formoreinformation,seetheCARESCAPEModularMonitorsSupplemental
InformationManual.
Audiblealarmsoffbehavior
DependingontheAudioAlarmdefaultsettingsconguredduringinstallation,you
canturnonorturnoffaudiblealarms.
136CARESCAPEModularMonitors2062971-001