ECG
3.Tapethestresslooptothepatient(excludingneonates).
Asecuredstresslooppreventsleadwirerotationabouttheelectrodesnap,
leadwiretuggingattheelectrode,andECGartifact.
3–leador5–leadECGelectrodeplacement
Fora3-leadwireelectrodeplacement,theR/RA,L/LA,andF/LLelectrodesshould
beused.
IEC
AAMI/AHA
Electrodeplacement
R(red)RA(white)Justbelowtherightclavicle.
L(yellow)LA(black)Justbelowtheleftclavicle.
UserdenedUserdenedForthe5-leadplacement,placethe
precordialelectrodeaccordingtothe
physician'spreference.
N(black)
RL(green)
Lowerrightedgeoftheribcage.
F(green)
LL(red)Lowerleftedgeoftheribcage.
6–leadECGelectrodeplacement
NOTE
PDMandPSMonly.
NOTE
For12RLmonitoring,a12RL12leadECGlicenseisrequired.
2062971-001CARESCAPEModularMonitors151