Airwaygases
UsingtheCARESCAPErespiratorymodulesand
compactairwaymodulesforO₂measurement
SelectingtheO
2
scale
IfthedifferencebetweenFiO
2
andEtO
2
ismorethan6%,changetheO
2
scale.
1.Selectagasrelatedparameterwindow.
2.SelecttheO2tab>Setup.
3.SelectanoptionfromtheScalelist.
SelectingtheO
2
sweepspeed
Thisselectionaffectsthewaveform.
1.Selectagasrelatedparameterwindow.
2.SelecttheO2tab>Setup.
3.SelectanoptionfromtheO2SweepSpeedlist.Theoptionsare0.625mm/s,
6.25mm/s,12.5mm/s,and50mm/s.
Thesmallerthevalue,theslowerthesweepspeed.
SettingO
2
alarms
1.Selectagasrelatedparameterwindow.
2.SelecttheO2tab>Alarms.
3.Checkthattherequiredalarm(EtO2orFiO2)ison,andsetitshighand/orlow
limitvalues.
UsingtheCARESCAPErespiratorymodules
andcompactairwaymodulesforAAandN₂O
measurement
Selectingtheagentscale
Everyanestheticagenthasitsowndefaultscalethatthemonitoruseswhendetecting
theagent.Youcanchangethescaleofanagentiftheamountusedishigherthan
thedefaultscale.DefaultscalesaregivenintheCARESCAPEModularMonitors
SupplementalInformationManual.
1.Selectagasrelatedparameterwindow.
2.SelecttheAgent/N2Otab>Setup.
3.SelectanoptionfromtheAgentScalelist.
Selectingtheagentsweepspeed
Thisselectionaffectsthewaveform.
294CARESCAPEModularMonitors2062971-001