Gasexchange
NOTE
PlaceallD-liteportsupwardswitha20°to45°tilttoprevent
condensedwaterfromenteringthesensorinteriorandthe
tubings.
NOTE
Whenmonitoringpediatricpatientswithtidalvolumesof15
to300ml,usethePedi-lite(+)sensor.Remembertoselectthe
sensortypeaccordingly.
Checkingthegasexchangemeasurement
1.Checkthatthewatertrapisempty.
2.OccludethesamplinglineandcheckthattheSamplelineblockedmessage
appearswithin30secondsandgaswaveformsareshowingzeroatthesametime.
Usingthegasexchangemeasurement
Selectingthegasexchangesensortype
Selectthesensortype(AdultorPedi)accordingtothesensoryouareusing.
1.Selectthegasexchangeparameterwindow.
2.SelecttheSetuptab.
3.SelectthesensortypefromtheSensorTypelist.
SelectingEEandRQaveragingtime
Averagedvaluesofenergyexpenditure(EE)andrespiratoryquotient(RQ)update
everyminute.ThebarintheEE+RQparameterwindowindicateswithgreencolorthe
amountofdatathatthemonitorusesforperformingtheaveragecalculations.
Whenapatientisadmitted/acaseisstarted,noaveragedvalueisdisplayeduntil10
minutesofvalidEEandRQdataisavailable.Afterthat,theEEandRQvaluesaregray
untilthereisenough(>1/5oftheselectedaveragingtime)datatoperformreliable
calculations.
1.Selectthegasexchangeparameterwindow.
2.SelecttheSetuptab.
3.SelectavaluefromtheEEAverageTimelist.
NOTE
Thevaluesintheparameterwindowturngraywhenever
thereisnotenoughvalidEEandRQaveragingdata.
NOTE
IftheRQisoutofthephysiologicalrange(<0.6or>1.3),the
monitordoesnotstoretheEEandRQvaluesinthetrend
historyorusethemforaveragecalculations.
WeightedVO
2
andVCO
2
TohaveweightedvaluesforVO
2
(VO2/kgorVO2/m2)andVCO
2
(VCO2/kgorVCO2/m2),
youmustenterthepatient’sheightandweight.ForVO2/kgandVCO2/kg,measured
valuesaredividedbythepatient’sweight,andforVO2/m2andVCO2/m2,measured
valuesaredividedbythepatient’sBSA.
342CARESCAPEModularMonitors2062971-001