Impedancerespiration
RespirationleadIIelectrodeplacement
ECGleadIIforchestorupperabdominalbreather
IEC
AAMI/AHA
Electrodeplacement
R(red)RA(white)Justbelowtherightclavicle.
F(green)
LL(red)Lowerleftedgeoftherib
cage.
RespirationleadRL-LLelectrodeplacement
NOTE
TheRL-LLrespirationleadisavailablewithPDMonly.
WhenmonitoringrespirationthroughtheRL-LLvector,useastandard5-or6-leadwire
electrodeplacement,exceptplacetheRLelectrodeonthefthintercostalspaceon
therightsideofthechest.ImpedancerespirationleadbetweenV5RandLLprovides
maximumrespirationsignalstrength,minimumnoise/artifactcontent,andminimum
cardiogenicartifact.
RL-LLvectorforabdominalbreather
IEC
AAMI/AHA
Electrodeplacement
N(black)
RL(green)
Fifthintercostalspaceonthe
right.
F(green)
LL(red)Lowerleftedgeoftherib
cage.