CO2withCAPNOSTATMainstream,CapnoFlexLF,andDualCO2modules
WARNING
CAPNOSTAT:Azerocellcalibrationisrequiredeachtimea
sensorisconnected.Failuretodosocanresultininaccurate
CO
2
values.
WARNING
CapnoFlexLF:Donotusethismoduleonpatientsthatcannot
toleratetheremovalof50ml/minfromtheirtotalminute
ventilation.
WARNING
DualCO2:Donotusethismoduleonpatientsthatcannot
toleratetheremovalof180ml/minfromtheirtotalminute
ventilation.
WARNING
CAPNOSTAT:Failuretoconguretheatmosphericpressure
correctlywillleadtoinaccurateCO
2
readingsinmmHgand
kPa.
CO
2
cautions
CAUTION
Donotapplypressurizedairorgastoanyoutletortubing
connectedtothemonitor.Pressuremaydestroysensitive
elements.
CO
2
pointstonote
Makesurethatyouareusingawatertrapthatiscompatiblewiththemodule:
●DualCO2:AquaKnot
●CAPNOSTATMainstream:nowatertrap
●CapnoFlexLFCO
2
:containedinsamplelines.
CO₂measurementsetup
EquipmentconnectionwithCAPNOSTATMainstream
module
WARNING
DonotplugtheCAPNOSTATMainstreamEtCO
2
sensorinto
theSolarCapnoFlexAdapter.Plugthesensordirectlyintothe
MainstreamEtCO
2
moduleortheDualCO2module.
2062971-001CARESCAPEModularMonitors309