EasyManua.ls Logo

GE CARESCAPE - CO2 cautions; CO2 points to note; CO₂ measurement setup; Equipment connection with CAPNOSTAT Mainstream module

GE CARESCAPE
548 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CO2withCAPNOSTATMainstream,CapnoFlexLF,andDualCO2modules
WARNING
CAPNOSTAT:Azerocellcalibrationisrequiredeachtimea
sensorisconnected.Failuretodosocanresultininaccurate
CO
2
values.
WARNING
CapnoFlexLF:Donotusethismoduleonpatientsthatcannot
toleratetheremovalof50ml/minfromtheirtotalminute
ventilation.
WARNING
DualCO2:Donotusethismoduleonpatientsthatcannot
toleratetheremovalof180ml/minfromtheirtotalminute
ventilation.
WARNING
CAPNOSTAT:Failuretoconguretheatmosphericpressure
correctlywillleadtoinaccurateCO
2
readingsinmmHgand
kPa.
CO
2
cautions
CAUTION
Donotapplypressurizedairorgastoanyoutletortubing
connectedtothemonitor.Pressuremaydestroysensitive
elements.
CO
2
pointstonote
Makesurethatyouareusingawatertrapthatiscompatiblewiththemodule:
DualCO2:AquaKnot
CAPNOSTATMainstream:nowatertrap
CapnoFlexLFCO
2
:containedinsamplelines.
CO₂measurementsetup
EquipmentconnectionwithCAPNOSTATMainstream
module
WARNING
DonotplugtheCAPNOSTATMainstreamEtCO
2
sensorinto
theSolarCapnoFlexAdapter.Plugthesensordirectlyintothe
MainstreamEtCO
2
moduleortheDualCO2module.
2062971-001CARESCAPEModularMonitors309