Invasivepressures
Displayedvalueswillswitchbetween:Art123/51( )andArt100/60( )
Invasivepressuretroubleshooting
ProblemSolution
ArtifactdetectedandtheSmartBPoptionisturned
on.
Checkthepatient.
Repositionthecatheter.
Zerothetransducer.
Ifproblempersists,turnofftheSmartBPoption.
IftheSmartBPoptionisturnedoff,usetheaudio
pausefeaturebeforedrawingbloodandbefore
zeroingtoreducenuisancealarms.
Invasivepressurereadingsseemunstable.Makesuretherearenoairbubblesinthe
transducersystems.
Flushandzero.
Placethetransduceronthepatient’sphlebostatic
axis.
---/---(80)Systolicanddiastolicpressurevaluesdo
notdisplay.
ThismaybeduetotheSmartBPoptiondetecting
artifact.Whenartifactsaredetected,onlyMean
valuesaredisplayed.
Checkthepatient.
TurnofftheSmartBPifrequired.
Invasivepressurewaveformisdisplayedbutno
numericvaluesaredisplayed.
Zerothechannel.Invasivepressurenumeric
valuesaredisplayedonlyforsuccessfullyzeroed
channels.
Zeroingofinvasivepressurechannel(s)fails.Ensurethatthechannelsareopentoair.
Thechannelstandbyisnotselectable,oritis
terminatedwithouttheusergivingtheactivation
request.
Ifthepressuresremainbetween10mmHgand
250mmHgfor10secondsormore,theselectionis
disabledorthestandbyisterminated.