Alarms
Formoreinformation,seetheCARESCAPEModularMonitorsSupplemental
InformationManual.
Whenaudiblealarmsareturnedoff:
●Allaudiblealarmsareturnedoffexceptforanyhighpriorityalarmscongured
tobreakthroughtheaudiooffsetting.
●Theaudiooffbellicondisplaysintheupperleftcornerofthedisplayscreen.
●Thealarmaudiopause/offareaofthealarmlightissolidbluewhenaudiblealarms
arepausedorwhenaudiooffisselectedforanalarmgroup.
Turningaudiblealarmson/off
Youcanturnon/offtheaudiblephysiologicalalarmtonesforanalarmgrouporfor
allalarms.
1.SelectAlarmSetupfromthemonitor’smainmenu.
2.SelecttheAudible&Visualtab.
3.Selectanalarmgroup.Choicesare:
●None:Noaudiblealarmsareturnedoff.
●ApneaAudioOff:Turnsoffaudiblealarmsforapnea,EtCO
2
,FiCO
2
,respiration
rate,Ppeaklow,PEEPe,PEEPtot,PEEPi,andMVexplimitalarms.
●ECGAudioOff:TurnsoffaudiblealarmsforallHRandPRsourcelimitand
arrhythmiaalarms.
●Apnea&ECGAudioOff:TurnsoffaudioalarmsforallHRandPRsourcelimit,
arrhythmia,apnea,EtCO
2
,FiCO
2
,respirationrate,Ppeaklow,PEEPe,PEEPtot,
PEEPi,andMVexplimitalarms.
●AllAlarmsAudioOff:Turnsoffallaudiblealarmsexceptsomehighpriority
alarmsdenedasbreakthroughalarms.
4.Toturnonallaudiblealarmsagain,selectActivateAllAudibleAlarms,orselect
Noneasinstructedabove.
NOTE
Ifalarmsareturnedoffforanyofthedenedalarmgroups
andanalarmoccurswithinthealarmgroup,abeeptonewill
soundevery2minutesasareminderthatalarmsareturned
off.
NOTE
Franceonly:Areminderbeeptonesoundsevery2minutes
whentheAudible&Visual>Nonesettingisnotselected.
Pauseaudiobehaviors
Selectingthepauseaudiokeyresultsindifferentalarmbehaviorsdependingon
whetherthealarmsareactiveand/orlatchedornot.Acknowledgingorpausingaudio
alarmsdoesnotaffectotheralarmindicators.Theywillstillcontinueindicatingalarms.
Whenthemonitorisonthenetwork,alarmscanalsobepausedandacknowledgedat
thecentralstation.
2062971-001CARESCAPEModularMonitors137