ECG
WARNING
—PACEMAKERINDICATION—Pacemakeractivityisindicated
ontheelectrocardiogramthroughthedisplayofadifferent
coloredpacemakermarkerpulse.Allpacemakermarker
pulsesappearuprightanduniformandshouldnotbeused
fordiagnosticinterpretation.
WARNING
—PATIENTHAZARD—Apacemakerpulsecanbecountedas
aQRSduringAsystolewhenpacemakerdetectionison.Keep
pacemakerpatientsundercloseobservation.
WARNING
—PATIENTHAZARD—Asystolemaynotbedetectedifthe
patienthasapacemakerthatproduceshigh-amplitudepacer
spikes,thepacemakerdetectionison,andaPDMisused.
Keeppacemakerpatientsundercloseobservation.
Pacemakerdetectionpointstonote
●PDM,TRAM,combinationmonitoring:Pacemakerdetectionmustbeturnedonat
themonitor.Itmustbeusedwheneverthemonitoredpatienthasapacemaker.
●E-modules:Pacemakerdetectionisalwayson.
●E-modules:Ifthepatienthasanatrialpacemaker,STcalculationscanbeperformed
ifthepacerspikedoesnotcoincidewiththeISOpoint’sadjustmentrange.
Selectingthepacemakerdetection
WithE-modulesthepacemakerdetectionisalwaysenabled.
WithPDM,TRAM,andcombinationmonitoringitmustbeturnedon.However,you
maydisablepacemakereventprocessingbyturningoffpacemakerdetection.When
pacemakerdetectionisturnedoff,themonitoringdeviceignorespacemakerpulse
detectionswhichmayadverselyaffecttheheartrateaccuracyofthemonitoring
device.
1.SelecttheHRparameterwindow.
2.SelecttheAdvancedtab.
3.SelectavaluefromthePacemakerDetectionlist.
Listoptionsareacquisitionmoduledependent:
●E-modules:
■Show:DisplayspacemakerspikesontheECGwaveform.
■Hide:HidesthepacemakerspikesontheECGwaveform.
■Sensitive:Increasespacemakerdetectionsensitivityanddisplaysthe
pacemakerspikesontheECGwaveform.Byselectingthisoptionyoucan
improvethedetectionofsmallamplitudepacemakers.However,thismodeis
alsomoresensitivetofalsepacemakerdetections.
●PDM:
■On:Turnsonpacemakerdetection.
■Off:Turnsoffpacemakerdetection.
●TRAM,combinationmonitoring:
■Off:Turnsoffpacemakerdetection.
2062971-001CARESCAPEModularMonitors173