ECG
SelectingtherstdisplayedECGlead
TheECG1LeadistherstECGleaddisplayedintheECGwaveformarea.Themonitor
usestheECG1Leadforsingle-leadanalysisifitisI,II,III,orV1.Ifitisanythingelse,
thenthefollowingmappingisused:V2toV6=V1,aVR=II,aVL=I,aVF=III.
1.SelecttheHRparameterwindow.
2.SelectaleadfromtheECG1Leadlist.
SelectingtheseconddisplayedECGlead
TheECG2LeadistheECGleaddisplayedaftertheECG1LeadintheECGwaveform
area.
1.SelecttheHRparameterwindow.
2.SelectaleadfromtheECG2Leadlist.
IfyourselectionisCascade,thedisplayedECG1Leadwaveformcontinuesinto
theECG2Leadwaveformarea.
SelectingthethirddisplayedECGlead
TheECG3LeadistheECGleaddisplayedaftertheECG2LeadintheECGwaveform
area.
1.SelecttheHRparameterwindow.
2.SelectaleadfromtheECG3Leadlist.
IfyourselectionisCascade,thedisplayedECG2Leadwaveformcontinuesinto
theECG3Leadwaveformarea.
SelectingtheVaECGlead
NOTE
12RLmonitoring-TheValeadistherstV-leadlabelused
witha6-leadwireECGcablefor12RLmonitoring.12RLis
possibleonlyiftheVaissettoV1.
TheVaLeadPositionselectionaffectstheSTnumerictrends.
Whenusinga6-leadwireECGcable,thefactorydefaultfortheValeadisV1,however
youmaychooseadifferentlead.
TheValeadistheonlyV-leadusedwitha5-leadwireECGcable.
TheValeadistheV-leaddatathatissenttoallremotedeviceslikethecentralstation.
1.SelecttheHRparameterwindow.
2.SelectaleadfromtheVaLeadPositionlist.
SelectingtheVbECGlead
NOTE
PDMandE-modulesonly.
NOTE
12RLmonitoring-TheVbleadisthesecondV-leadlabelused
witha6-leadwireECGcablefor12RLmonitoringandmust
besettoV5.
Thisselectionisnotavailablewithcombinationmonitoring.