Viewingothermonitoredpatients
6.Selectwhichalarmprioritylevelsyouwantnoticationof:
High:Opensabed-to-bedwindowforremotepatientsinahighalarmpriority
condition.
High,Med:Opensabed-to-bedwindowforremotepatientsinahighor
mediumalarmprioritycondition.
High,Med,Low:Opensabed-to-bedwindowforremotepatientsinahigh,
medium,orlowalarmprioritycondition.
Changingthesettingsformultiplebeds
Youcanselectacareunitandchangeallofthelistedremotepatientbedstoasingle
noticationsettingand/orasinglealarmprioritysetting.Iftherearemorethan40
bedsintheunit,thesettingswillbechangedfortherst40bedsonly.
1.SelectData&Pages.
2.SelectOtherPatients.
3.SelecttheReceiveAlarmstab.
4.SelectacareunitfromtheUnitlist.
5.SelectasettingfromtheChangeAllNoticationslist:
Off:Remotealarmnoticationisturnedoff.
Message:Remotealarmmessagesdisplayinthealarmarea.
AutoView:Bed-to-bedwindowopensimmediatelyifnootherprocedureor
setupwindowiscurrentlyopen.Otherwise,aremotealarmmessagedisplays
inthealarmmessagearea.
AutoViewAlways:Immediatelyclosesanyopenprocedureorsetupwindows
andopensabed-to-bedwindow.
6.SelectasettingfromtheChangeAllPrioritieslist:High;High,Med;High,Med,
Low.
Nextalarmingremotebedtoscreen
Ifyouhaveabed-to-bedwindowopenandanotherAutoVieworAutoViewAlways
bedgoesintoalarm,theselectionViewNextPatientbecomesselectable.Itallows
youtoopenthebed-to-bedwindowtoviewthenexthighestandnewestalarming
patientbed.
Viewingremotepatientbeds
Youcanselectandviewanetworkedalarmingornon-alarmingremotepatientbed.
1.SelectData&Pages.
2.SelectOtherPatients.
3.SelecttheViewPatientstab.
4.SelectacareunitfromtheUnitlist.
Alistofremotepatientbedsisdisplayedfortheselectedcareunit.
2062971-001CARESCAPEModularMonitors445