Non-invasivebloodpressure
3.StopthemeasurementbyselectingNIBPAuto>StopCycling.
StartingorstoppingtheNIBPAutofromthemonitor’smainmenu
1.SelectNIBPAutoStart.
2.StopthemeasurementbyselectingNIBPAutoStop.
StartingorstoppingtheNIBPAutowiththePSMmodulekey
1.PresstheAutoOn/Offkey.
2.StopthemeasurementbypressingtheAutoOn/Offkeyagain.
STATmode
NOTE
NotavailableintheNICUsoftwarepackage.
TheSTATmodeinitiatesacontinuouscycleofmeasurementsforveminutes.The
messageSTATdisplaysintheNIBPparameterwindowwhenSTATisstarted.Anew
NIBPmeasurementstartsafterthepreviousmeasurementcompletes.Theamountof
timebetweenmeasurementsvaries.ForPDMandPSM,itisatleastfoursecondsfor
adultandchildandatleasteightsecondsforinfant.ForTRAM,thedelaybetween
measurementsistwoseconds.Theearlysystolicvalueismeasuredanddisplayed
untilthenalresultisavailable.Afterveminutes,themonitorautomaticallyreturns
tothepreviouslyselectedcyclingintervalormanualmode.
StartingorstoppingaStatNIBPmeasurement
YoucansettheNIBPmeasurementtocontinueforveconsecutiveminutes.
1.SelecttheNIBPparameterwindow.
2.SelectStartStat.
3.StopthemeasurementbyselectingStopStat.
Venousstasis
NOTE
PSMonly.
Thevenousstasismodeinitiatesinationandholdsaconstantpressureinthecuff
tohelpvenouscannulation.ThemessageStasisdisplaysintheNIBPparameter
windowwhenvenousstasismodeisstarted.Duringthelast10seconds,theStasis
messagebeginstoashtoindicatethemonitorisabouttoreturntothepreviously
selectedcyclingintervalormanualmode.Stasismeasurementpressureiscontrolled
bythePSMinternally.
VenousstasisallowsyoutoapplycontinuousNIBPcuffpressureforashortperiodof
time.Thecuffinationpressureanddurationaredependentonthedetectedcuffor
selectedinationlimits.
Cuff
Pressure
Duration
Adult80±5mmHg(10.7±0.7kPa)
2minutes
Child60±5mmHg(8.0±0.7kPa)
2minutes
Infant40±5mmHg(5.3±0.7kPa)
1minute