ECG
STdepressionSTelevation
Enablingischemicburden
Thisselectionisnotavailablewithcombinationmonitoring.
1.SelecttheSTparameterwindow.
2.SelectTrendView.
3.SelectIschemicBurden.
4.SelectthecheckboxforIschemicBurden.
Settingtheischemicburdenlimits
Thisselectionisnotavailablewithcombinationmonitoring.
Youcansetthelowerandupperthresholdvalues.
1.SelecttheSTparameterwindow.
2.SelectTrendView.
3.SelectIschemicBurden.
4.SetthelowerthresholdvaluewiththeDepressionLimit(mm)arrows.
5.SettheupperthresholdvaluewiththeElevationLimit(mm)arrows.
STalarmlimits
DependingonwhathasbeenselectedintheCareUnitSettingsforSTalarms,you
maysettheSTalarmlimitsforaleadgroup,forindividualleads,orforallleadsrelative
tothepatient’scurrentmeasurements.
Theleadsassociatedwitheachleadgroupareasfollows:
Anterior:V1,V2/dV2,V3/dV3,V4/dV4.
Inferior:II,III,aVF.
Lateral:V5,V6/dV6,I,aVL.
ThedindV2,dV3,dV4,anddV6aboverepresentsthederivedleadvalue.
Settingalarmlimitsforleadgroups
1.SelecttheSTparameterwindow.
2.SelectAlarms.
3.SelectAlarmOnforanECGleadgroup:Anterior,Inferior,orLateral.
Ifthealarmislocked,thereisalocksymbolbesidetheselectionandtheselection
isnotavailable.
4.Setupperandloweralarmlimitswiththearrows.