Entropy
SelectingtheEntropytrendlength
ThissettingaffectsthewidthoftheEntropymicrotrendintheparameterwindow.
1.Pressthe
modulekey,orselecttheEntropyparameterwindow.
2.SelectSetup.
3.SelectatrendlengthfromtheTrendLengthlist.
UsingthemanualEntropysensorcheck
Wheneverrequired,youcanperformthesensorcheckmanually.Pressthe
modulekey,or:
1.Pressthe
modulekey,orselecttheEntropyparameterwindow.
2.SelectSetup>CheckSensor.
3.Observetheresultsonthescreen.Donotpressthesensorduringthecheckto
avoidsignalnoise.
Themeasurementcontinuesautomaticallyafterthesensorhaspassedthecheck.
UsingtheautomaticEntropysensorcheck
Sensorimpedancecheckisperformedatthestartupwheneverasensororan
electrodeisattached.Iftheautomaticsensorcheckhasbeenselectedon,thecheck
isalsoperformedperiodicallyevery10minutes.
1.Pressthemodulekey,orselecttheEntropyparameterwindow.
2.SelectSetup.
3.SelectCheckSensor>Automatic.
BypassingtheEntropysensorcheck
Ifthesensordoesnotpasstheimpedancecheck,thisoptionbecomesselectable.It
allowsyoutostartthemeasurementwithoutcompletingthesensorcheck.Inthis
case,themeasurementmaybeunreliable.
1.Pressthe
modulekey,orselecttheEntropyparameterwindow.
2.SelectSetup.
3.SelectBypassCheck.
SettingEntropyalarmlimits
Youcansetthelimitalarmsonoroffandadjusttheiractivationlimitsaccording
toyourneeds.
1.Pressthemodulekey,orselecttheEntropyparameterwindow.
2.SelectSetup.