Temperature
Temperaturechannel
Measurementsource,B850Measurementsource,B650
orB450
T4
E-PTmoduleE-PTmodule
TbloodPDM(Masimo),PDM(Nellcor),
TRAM451,TRAM451N5,
TRAM451M,TRAM851,TRAM
851N5,TRAM851M,E-COP,
orE-COPSvmoduleorUnity
ID.
PDM(Masimo),PDM(Nellcor),
E-COP,orE-COPSvmoduleor
UnityID.
Startingthetemperaturemeasurement
Connectthetemperatureprobetostartthemeasurement.Iftheparameterwindow
displaysOffinthevalueeld:
1.Selectthetemperatureparameterwindow.
2.ConrmthatMeasurement>Onisselected.
Changingthetemperaturesitelabel
1.Selectthetemperatureparameterwindow.
2.ChooseasitelabelfromtheLabellist.
NOTE
Inadditiontothedefaultchannellabels,youcancongureup
tosixuser-denedchannellabelsinthecareunitsettings.See
theCARESCAPEModularMonitorsSupplementalInformation
Manualfordetails.
Displayingthedeltavaluebetweentwotemperature
channels
NOTE
Thisselectionisavailablewhentwotemperaturesare
displayedinthesametemperatureparameterwindow.
1.Selectthetemperatureparameterwindow.
2.SelectShowTx-Ty(e.g.,T2-T1).
Settingtemperaturealarms
1.Selectthetemperatureparameterwindow.
2.SelectAlarms.
3.Chooseatemperaturechannelortemperaturedelta.
NOTE
Ifthefeatureisnotactive,thealarmlimitsaregreyedout.
SelectAlarmOntosetthealarms.
4.Setthealarmlimits.
NOTE
Thehighlimitalarmsettingfordeltavaluesisalsoadjustable.
NOTE
Ifthesettingofanalarmlimithasbeendisabledduring
conguration,thesettingismarkedwithalocksymbol.
258CARESCAPEModularMonitors2062971-001