Non-invasivebloodpressure
2.StopthemeasurementbypressingtheStartCancelkeyagain.
AutomaticNIBPmeasurements
AutomaticNIBPmeasurementsandmonitorclock
synchronization
NOTE
TRAMmodulesdonotsupportautomaticclock
synchronization.
Clocksynchronizationtiming(cyclingsync)automaticallysynchronizesyourautomatic
NIBPmeasurementtimeintervalswiththemonitorclock.Forexample,ifautomatic
measurementsareinitiatedforveminuteintervalsat4:02,therstmeasurementis
takenimmediatelyat4:02.Thenextmeasurementwillbetakenat4:05(intervaland
clockarenowsynchronized).Allmeasurementswillcontinueatveminuteintervals
(i.e.,4:10,4:15,etc.)
Therewillalwaysbeatleasta30seconddelaybetweentwoconsecutiveNIBP
measurementsduringautocycling.Ifanautomaticmeasurementcompleteswith
lessthan30secondstothenextscheduledmeasurement,themonitorwilldelaythe
scheduledmeasurementuntil30secondshavepassed.Thecyclingsynchronization
isnotdoneduringthese30secondsbutitwillbedoneafterthedelayedauto
measurementstarts.
Exampleswitha5minutecyclingtime:
Completionwithlessthan30secondsto
thenextscheduledmeasurement
Completionafterthenextscheduled
measurementshouldhavestarted
Firstautomeasurementstarts:4:59:00
Firstautomeasurementcompletes:
4:59:40
Secondautomeasurementstarts:5:00:20
(notclocksynchronized)
Thirdautomeasurementstarts:5:05:00
(clocksynchronized)
Firstautomeasurementstarts:4:59:00
Firstautomeasurementcompletes:
5:00:10
Secondautomeasurementstarts:5:05:00
SettingthecycletimebetweenNIBPmeasurements
ToautomaticallymeasureNIBPatsettimeintervals,youmustrstsetthecycletime.
1.SelecttheNIBPparameterwindow.
2.SelectthecycletimefromtheCycleTimelist.
NIBPAutomode
TheNIBPAutomodeinitiatesrepeatedmeasurementsfortheselectedCycleTime.
Therewillbeatleasta30seconddelaybetweentwoconsecutiveNIBPmeasurements
duringautocycling.
StartingorstoppingtheNIBPAutofromtheNIBPSetupmenu
1.SelecttheNIBPparameterwindow.
2.SelectStartCyclingforNIBPAuto.
224CARESCAPEModularMonitors2062971-001