Bispectralindex
CAUTION
WhenconnectingordisconnectingBIS,takecarenottotouch
theexposedcontactsofeitherconnector.Damagedueto
electrostaticdischargemayresulttotheequipment.
BISindicationsforuse
TheBISmoduleisintendedforusebypersonneltrainedinitsproperuse.Itisintended
foruseonadultandpediatricpatientswithinahospitalormedicalfacilityproviding
patientcaretomonitorthestateofthebrainbydataacquisitionofEEGsignals.
TheBispectralindex(BIS),aprocessedEEGvariable,andonecomponentoftheBIS
module,maybeusedinadultsasanaidinmonitoringtheeffectsofcertainanesthetic
agents.TheBispectralindexisacomplextechnology,intendedforuseonlyasan
adjuncttoclinicaljudgementandtraining.Inaddition,theclinicalutility,risk/benet,
andapplicationofBIShavenotundergonefullevaluationinthepediatricpopulation.
BISmeasurementlimitations
ThismeasurementisnotavailableintheNICUsoftwarepackage.
E-modulesusedforthismeasurementarenotsuitableforusewithneonatal
patients.
BISisacomplexmonitoringtechnologyintendedforuseonlyasanadjunctto
clinicaljudgmentandtraining.Clinicaljudgmentshouldalwaysbeusedwhen
interpretingBISinconjunctionwithotheravailableclinicalsigns.RelianceonBIS
aloneforintraoperativeanestheticmanagementisnotrecommended.
BISvaluesshouldbeinterpretedcautiouslywithcertainanestheticcombinations,
suchasthoserelyingprimarilyoneitherketamineornitrousoxide/narcoticsto
produceunconsciousness.
Externalradiatingdevicesmaydisturbthemeasurement.
PoorsignalqualitymayleadtoinappropriateBISvalues.
ArtifactmayleadtoinappropriateBISvalues.Potentialartifactmaybecausedby
unusualorexcessiveelectricalinterferenceorhighEMGactivitylikeshivering,
muscleactivityorrigidity,sustainedeyemovements,headandbodymotion.Also,
impropersensorplacementandpoorskincontact(highimpedance)maycause
artifactandinterferewiththemeasurement.
Duetolimitedclinicalexperienceinthefollowingapplications,theBISvalueshould
beinterpretedcautiouslyinpatientswithknownneurologicaldisordersandthose
takingpsychoactivemedications.
BISpointstonote
Makesurethatthesensorconnectorsofthepatientinterfacecablearenotin
contactwithuids.
BISsensorsaredisposable,forsinglepatientuseonly,andnotmadewithnatural
rubberlatex.
Checkthesensorexpirationdateonthesensorpackage.
Donotuseasensorformorethan24hours.
UseonlyCovidienBISsensors.