Messages
Messagesrelatedtoimpedancerespiration
measurement
Forinformationregardingalarmprioritiesandescalationtimes,seetheCARESCAPEModular
MonitorsSupplementalInformationManual.
Makesureyouarefamiliarwiththegenericlayoutofthescreen.Thiswillhelpyouidentify
whereonscreenthefollowingmessagesappear.Themessagelocationisindicatedwiththe
followingabbreviations:
al.area=alarmarea
param.=parameterwindow
report=reportview
wavef.=waveformarea
Message
Location
ExplanationWhattodo
Apnea(Imped)al.area,
wavef.
Nobreathingdetected.Checkthepatientstatus.
Checktheventilatorand
breathingstatus.
Apneadeactivated
param.
Thecasehasjustbeen
started/patientadmittedonthe
monitor,orthemeasurement
hasjustbeenstarted.
Wait.Themessage
disappearsafterthemonitor
detects3breaths(E-modules)
ortherespirationrateis≥3
(PDM,TRAM).
Artifact
param.,
wavef.
Cardiacartifacthasbeen
detected.
Checkthepatientstatus.
Selectalternateleadsto
monitor.
Increasesensitivitysettings.
Selectalternatelead
placement.
Relearnrespiration.
Cardiacartifactal.areaCardiacartifacthasbeen
detected.
Checkthepatientstatus.
Selectalternateleadsto
monitor.
Increasesensitivitysettings.
Selectalternatelead
placement.
Relearnrespiration.
LA/Lleadoff
LeadIfailed
LeadIIfailed
LeadRL-LLfailed
LL/Fleadoff
RA/Rleadoff
param.,
wavef.
Oneoftheelectrodesisoff.Checktheelectrodesand
theirconnections.
468CARESCAPEModularMonitors2062971-001