ExplanationWhattodo
SVTachyal.area,
wavef.
Physiologicalalarm.Checkthepatientstatus.
Telemetrypatient:No
12SLavailable
param.
12SLisnotavailablewith
telemetry.
If12SLisrequired,use
amodulethatisdirectly
connectedtothemonitor
insteadofatelemetry
transmitter.
Tachyal.area,
wavef.
Physiologicalalarm.Checkthepatientstatus.
Telemetrybattery
empty
al.areaThetelemetrytransmitter’s
batteryhasnochargeleft.
Changethebattery.
Trigeminy
al.area,
wavef.
Physiologicalalarm.Checkthepatientstatus.
TTXOffNetworkal.areaNoparameterorwaveform
datahasbeenreceivedfromthe
telemetrytransmitterformore
than5seconds.
Waittoseeifthecondition
resolves.Iftheproblem
persists,contactauthorized
servicepersonnel.
VBradyal.area,
wavef.
Physiologicalalarm.Checkthepatientstatus.
VFib/VTachal.area,
param.,
wavef.
Physiologicalalarm.Checkthepatientstatus.
VleadAisnotV1orV
leadBisnotV5
report
ValeadisnotV1and/orVblead
isnotV5.
Checkthesettingsand
correctifnecessary.
Vleadsoff
report
AllVleadsareoff.ConnecttheVleadstothe
patient.
VTachal.area,
param.,
wavef.
Physiologicalalarm.Checkthepatientstatus.
V2/C2leadoff
V3/C3leadoff
V4/C4leadoff
V5/C5leadoff
V6/C6leadoff
V/Cleadoff
Va/Caleadoff
Vb/Cbleadoff
wavef.Anelectrodeisdisconnected.Checktheelectrodes.
VT>2
al.area,
wavef.
Physiologicalalarm.Checkthepatientstatus.