Messages
Message
Location
ExplanationWhattodo
●SVTachy●al.area,
wavef.
Physiologicalalarm.●Checkthepatientstatus.
●Telemetrypatient:No
12SLavailable
●param.
12SLisnotavailablewith
telemetry.
●If12SLisrequired,use
amodulethatisdirectly
connectedtothemonitor
insteadofatelemetry
transmitter.
●Tachy●al.area,
wavef.
Physiologicalalarm.●Checkthepatientstatus.
●Telemetrybattery
empty
●al.areaThetelemetrytransmitter’s
batteryhasnochargeleft.
●Changethebattery.
●Trigeminy
●al.area,
wavef.
Physiologicalalarm.●Checkthepatientstatus.
●TTXOffNetwork●al.areaNoparameterorwaveform
datahasbeenreceivedfromthe
telemetrytransmitterformore
than5seconds.
●Waittoseeifthecondition
resolves.Iftheproblem
persists,contactauthorized
servicepersonnel.
●VBrady●al.area,
wavef.
Physiologicalalarm.●Checkthepatientstatus.
●VFib/VTach●al.area,
param.,
wavef.
Physiologicalalarm.●Checkthepatientstatus.
●VleadAisnotV1orV
leadBisnotV5
●report
ValeadisnotV1and/orVblead
isnotV5.
●Checkthesettingsand
correctifnecessary.
●Vleadsoff
●report
AllVleadsareoff.●ConnecttheVleadstothe
patient.
●VTach●al.area,
param.,
wavef.
Physiologicalalarm.●Checkthepatientstatus.
●V2/C2leadoff
●V3/C3leadoff
●V4/C4leadoff
●V5/C5leadoff
●V6/C6leadoff
●V/Cleadoff
●Va/Caleadoff
●Vb/Cbleadoff
●wavef.Anelectrodeisdisconnected.●Checktheelectrodes.
●VT>2
●al.area,
wavef.
Physiologicalalarm.●Checkthepatientstatus.
2062971-001CARESCAPEModularMonitors467