ECG
WARNING
ELECTRODESWheneverpatientdebrillationis
apossibility,usenon-polarizing(silver/silverchloride
construction)electrodesforECGmonitoring.Polarizing
electrodes(stainlesssteelorsilverconstructed)maycause
theelectrodestoretainaresidualchargeafterdebrillation.A
residualchargewillblockacquisitionoftheECGsignal.
WARNING
DEFIBRILLATORPRECAUTIONSPatientsignalinputs
labeledwiththeCFandBFsymbolswithpaddlesare
protectedagainstdamageresultingfromdebrillation
voltages.Toensureproperdebrillatorprotection,useonly
therecommendedcablesandleadwires.
WARNING
HEARTRATEALARMINTERFERENCEPoorcablepositioning
orimproperelectrodepreparationmaycauselineisolation
monitortransientstoresembleactualcardiacwaveformsand
thusinhibitheartratealarms.Tominimizethisproblem,follow
properelectrodeplacementandcablepositioningguidelines
providedwiththisproduct.
ECGcautions
CAUTION
Thepatient'sskinmaybecomeirritatedafterprolonged
contactwithelectrodegeloradhesive.
CAUTION
Toassureaccurate12leadanalysiswhenusinga10-leadwire
patientcable,youmustverifythatthecorrectleadwireblockis
pluggedintotheappropriatesideofthecable.TheV2through
V6leadwireblockiscolor-codedbrown(AHA)orwhite(IEC).
ECGmeasurementlimitations
E-modulesusedforthismeasurementarenotsuitableforusewithneonatal
patients.
ThemonitorwilldisplayaLeadsoffmessageinaninputoverloadcondition,or
upondisconnectionofelectrodeleadwires.
ECGpointstonote
Pre-gelledECGelectrodesarerecommended.Checktheexpirationdate.
Makesuretheelectrodegelismoist.
Makesuretheelectrodeshavegoodskincontact.
Replaceallelectrodesatleastevery24to48hours.
UsetheMulti-Linkelectrosurgicalunit(ESU)ECGpatientcablewhenusingthe
monitorinthepresenceofanelectrosurgicalunit.Thiscable,withabuilt-inESU
lter,helpsreduceelectrosurgicalnoisedetectedontheECGsignal.
E-modules:Wheneveracable,electrodeorV-leadischanged,themonitor
automaticallyrelearns.
DependingontheECGmoduleused,notallECGmeasurementsandsettingsare
availabletovieworchange.
SelecttheDeactivateECGLeadsOffoptiontoremoveaLeadsoffmessagefrom
thedisplaywhenacableisoff.
2062971-001CARESCAPEModularMonitors149