ECG
WARNING
Thisdeviceusesacomputerized12leadECGanalysis
program,whichcanbeusedasatoolingeneratingECG
recordsthatprovideECGmeasurementsandinterpretative
statementsfromtheECGrecordings.Theinterpretive
statementsareonlysignicantwhenusedinconjunction
withclinicalndings.AllECGrecordsshouldbeoverreadby
aqualiedphysician.Toensureaccuracy,useonlytheECG
recordsforphysicianinterpretation.
WARNING
Whentransitioningfroma10-leadcabletoa5-leadcable,
selecttheUpdateLeadSetoptiontocleartheLeadoff
messagefromthedisplay.
WARNING
Disconnectedelectrodesorlooseelectrodeconnectionscan
leadtomissedcriticalseverityECGalarms.Ifthemonitor
reportsLeadsoffafterselectingUpdateLeadSetoption,
alwayschecktheelectrodeconnectionstothepatient.
WARNING
—CONDUCTIVECONNECTIONS—Extremecaremustbe
exercisedwhenapplyingmedicalelectricalequipment.Many
partsofthehuman/machinecircuitareconductive,such
asthepatient,connectors,electrodes,transducers.Itis
veryimportantthattheseconductivepartsdonotcome
intocontactwithothergrounded,conductivepartswhen
connectedtotheisolatedpatientinputofthedevice.Such
contactwouldbridgethepatient’sisolationandcancelthe
protectionprovidedbytheisolatedinput.
WARNING
—DELAYEDASYSTOLEALARM—Thepulsatileheartratemay
haveaslowerresponsetimethantheelectricalheartrate
wherethereisalowperfusionpatientcondition.Whenusing
theIntelliRatefeatureinthissituation,themonitormaydelay
callinganASYSTOLEpatientalarm.Theusermayelecttoturn
theIntelliRatefeatureoffforpatientsatriskoftheseevents,
otherwisepatienttreatmentmaybedelayed.Suchpatients
shouldalwaysbekeptundercloseobservation.
WARNING
—NOISYECGalarm—TheNoisyECGalarmindicatesthatthe
systemisnolongermonitoringECGandtheremaybenoHR
high,HRlow,TachyorBradyalarms.Ifyouadjustthealarm
prioritylevellowerthanthedefaultvalue,keepthepatient
underclosesurveillance.
WARNING
—INACCURATEHEARTRATEINDICATION—PDM—The
electricalandpulsatileheartratevaluesprovidedbythe
variousmonitoredparameters(ECG,SpO
2
,bloodpressures)
maydiffermarkedly.Thesedifferencesmaybedueto
underlyingphysiologicconditions(e.g.,electromechanical
dissociation,pulselesselectricalactivity,non-perfusing
rhythms)ortoinaccuraciesintheheartratevaluescaused
byartifactorpoorsignalquality.Theusermayelecttoturn
theIntelliRatefeatureoffforpatientsatriskoftheseevents,
otherwisepatienttreatmentmaybedelayed.Suchpatients
shouldalwaysbekeptundercloseobservation.
148CARESCAPEModularMonitors2062971-001